PURPOSE: To compare ablation zones created with equal amounts of 2.45 GHz microwave and 480 kHz radiofrequency (RF) energy in ex vivo liver and lung. METHODS: A total of 38 ablations were performed in ex vivo liver and lung for 10 min each. Nineteen RF ablations (nine liver, ten lung) were performed with a 480 kHz system (200 W max, impedance-based pulsing) and cooled electrode while measuring the average RF power applied. Nineteen microwave ablations (nine liver, ten lung) were then created using a cooled triaxial antenna to deliver 2.45 GHz at the same power level as in RF experiments. Ablation zones were then sectioned and measured for minimum, maximum and mean diameters, and circularity. Measurements were compared using t-tests, with P < 0.05 indicating statistical significance. RESULTS: Mean diameters of microwave ablations were greater than RF ablations in both liver and lung (4.4 +/- 0.3 vs 3.3 +/- 0.2 cm in liver; 2.45 +/- 0.3 vs 1.6 +/- 0.5 cm in lungs; P < 0.0005 all comparisons). There was no significant difference in the mean power applied during microwave or RF ablations in either organ (54.44 +/- 1.71 W vs 56.4 +/- 6.7 W in liver, P > 0.05; 40 +/- 0.95 W vs 44.9 +/- 7.1 W in lung, P > 0.05). CONCLUSIONS: Using a single cooled applicator, microwave energy at 2.45 GHz produces larger ablations than an equivalent amount of 480 kHz RF energy in normal liver and lung. This was more apparent in lung, likely due to the high baseline impedance which limits RF, but not microwave power delivery.
PURPOSE: To compare ablation zones created with equal amounts of 2.45 GHz microwave and 480 kHz radiofrequency (RF) energy in ex vivo liver and lung. METHODS: A total of 38 ablations were performed in ex vivo liver and lung for 10 min each. Nineteen RF ablations (nine liver, ten lung) were performed with a 480 kHz system (200 W max, impedance-based pulsing) and cooled electrode while measuring the average RF power applied. Nineteen microwave ablations (nine liver, ten lung) were then created using a cooled triaxial antenna to deliver 2.45 GHz at the same power level as in RF experiments. Ablation zones were then sectioned and measured for minimum, maximum and mean diameters, and circularity. Measurements were compared using t-tests, with P < 0.05 indicating statistical significance. RESULTS: Mean diameters of microwave ablations were greater than RF ablations in both liver and lung (4.4 +/- 0.3 vs 3.3 +/- 0.2 cm in liver; 2.45 +/- 0.3 vs 1.6 +/- 0.5 cm in lungs; P < 0.0005 all comparisons). There was no significant difference in the mean power applied during microwave or RF ablations in either organ (54.44 +/- 1.71 W vs 56.4 +/- 6.7 W in liver, P > 0.05; 40 +/- 0.95 W vs 44.9 +/- 7.1 W in lung, P > 0.05). CONCLUSIONS: Using a single cooled applicator, microwave energy at 2.45 GHz produces larger ablations than an equivalent amount of 480 kHz RF energy in normal liver and lung. This was more apparent in lung, likely due to the high baseline impedance which limits RF, but not microwave power delivery.
Authors: Stephanie A Solazzo; Zhengjun Liu; S Melvyn Lobo; Muneeb Ahmed; Andrew U Hines-Peralta; Robert E Lenkinski; S Nahum Goldberg Journal: Radiology Date: 2005-08 Impact factor: 11.105
Authors: Nam C Yu; Steven S Raman; Young Jun Kim; Charles Lassman; Xinlian Chang; David S K Lu Journal: J Vasc Interv Radiol Date: 2008-05-27 Impact factor: 3.464
Authors: David S K Lu; Steven S Raman; Piyaporn Limanond; Donya Aziz; James Economou; Ronald Busuttil; James Sayre Journal: J Vasc Interv Radiol Date: 2003-10 Impact factor: 3.464
Authors: Caroline J Simon; Damian E Dupuy; Thomas A DiPetrillo; Howard P Safran; C Alexander Grieco; Thomas Ng; William W Mayo-Smith Journal: Radiology Date: 2007-04 Impact factor: 11.105
Authors: David A Iannitti; Robert C G Martin; Caroline J Simon; William W Hope; William L Newcomb; Kelly M McMasters; Damian Dupuy Journal: HPB (Oxford) Date: 2007 Impact factor: 3.647
Authors: Farrah J Wolf; David J Grand; Jason T Machan; Thomas A Dipetrillo; William W Mayo-Smith; Damian E Dupuy Journal: Radiology Date: 2008-03-27 Impact factor: 11.105
Authors: Stefan Stättner; Florian Primavesi; Vincent S Yip; Robert P Jones; Dietmar Öfner; Hassan Z Malik; Stephen W Fenwick; Graeme J Poston Journal: Surg Today Date: 2014-03-16 Impact factor: 2.549
Authors: Maria Franca Meloni; Jason Chiang; Paul F Laeseke; Christoph F Dietrich; Angela Sannino; Marco Solbiati; Elisabetta Nocerino; Christopher L Brace; Fred T Lee Journal: Int J Hyperthermia Date: 2016-08-02 Impact factor: 3.914
Authors: Chunxia Su; Mathias Meyer; Robert Pirker; Wieland Voigt; Jingyun Shi; Lothar Pilz; Rudolf M Huber; Yilong Wu; Jinghong Wang; Yonglan He; Xuan Wang; Jian Zhang; Xiuyi Zhi; Meiqi Shi; Bo Zhu; Stefan S Schoenberg; Thomas Henzler; Christian Manegold; Caicun Zhou; Eric Dominic Roessner Journal: Transl Lung Cancer Res Date: 2016-08