Literature DB >> 20632595

MapCHECK used for rotational IMRT measurements: step-and-shoot, TomoTherapy, RapidArc.

Paul A Jursinic1, Renu Sharma, Jim Reuter.   

Abstract

PURPOSE: To measure patient-specific-QA dose distributions with a 2D array of diodes, the MapCHECK, for the dose delivered, with step-and-shoot and rotational IMRT.
METHODS: Two MapCHECKs were used that had different styles of diode connection. These MapCHECKs were used in their original manufactured configuration and in a modified configuration. The modification made in the clinic consists of filling air gaps with sheets of Lucite that had custom-machined slots for the diodes and by adding pieces of copper to offset the intrinsic asymmetry of the diodes. The MapCHECKs were housed in a tight-fitting phantom fabricated from solid water. Measurements were made on IMRT treatment plans delivered with Varian linear accelerators with step-and-shoot and RapidArc methods, and a TomoTherapy machine with helical delivery. Patient plans and QA plans were developed with the XiO, Eclipse, and TomoTherapy planning systems. All MapCHECK data were analyzed with its commercially available software.
RESULTS: Kilovoltage CT imaging of the MapCHECK in its phantom has streak artifacts from high atomic number components. These artifacts must be corrected in order to obtain accurate calculation of dose. The original MapCHECK is found to have an angular dependence of +/- 20%. A modification of the MapCHECK has been made that reduces the angular dependence to +/- 2%. Proper compensation for attenuation by the treatment couch is also necessary for accurate results. The modified MapCHECK has been successfully used for doing patient-specific QA for IMRT treatments delivered with step-and-shoot, TomoTherapy, and RapidArc methods. Treatment plans that require large amounts of fluence from the 90 degrees and 270 degrees directions have significantly better QA measurements when made with modified MapCHECKs.
CONCLUSIONS: When properly modified, the MapCHECK response becomes isotropic. It can then be used for patient-specific QA measurements for IMRT treatments delivered with step-and-shoot and rotational techniques such as helical Tomotherapy and RapidArc.

Entities:  

Mesh:

Year:  2010        PMID: 20632595     DOI: 10.1118/1.3431994

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  20 in total

1.  Error Detectability of Isodose Volumes as ROIs in Prostate Intensity-modulated RT QA.

Authors:  Ryuta Nakahara; Masayuki Fujiwara; Haruyuki Takaki; Masao Tanooka; Kentaro Ishii; Ryu Kawamorita; Koichiro Yamakado
Journal:  In Vivo       Date:  2022 Jul-Aug       Impact factor: 2.406

2.  Comparison of Dosimetric Performance among Commercial Quality Assurance Systems for Verifying Pretreatment Plans of Stereotactic Body Radiotherapy Using Flattening-Filter-Free Beams.

Authors:  Jin Beom Chung; Sang Won Kang; Keun Yong Eom; Changhoon Song; Kyoung Sik Choi; Tae Suk Suh
Journal:  J Korean Med Sci       Date:  2016-11       Impact factor: 2.153

3.  Anatomy-based, patient-specific VMAT QA using EPID or MLC log files.

Authors:  Dewayne L Defoor; Luis A Vazquez-Quino; Panayiotis Mavroidis; Nikos Papanikolaou; Sotirios Stathakis
Journal:  J Appl Clin Med Phys       Date:  2015-05-08       Impact factor: 2.102

4.  A comprehensive comparison study of three different planar IMRT QA techniques using MapCHECK 2.

Authors:  Vance P Keeling; Salahuddin Ahmad; Hosang Jin
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

5.  Evaluation of dosimetric effect caused by slowing with multi-leaf collimator (MLC) leaves for volumetric modulated arc therapy (VMAT).

Authors:  Zhengzheng Xu; Iris Z Wang; Lalith K Kumaraswamy; Matthew B Podgorsak
Journal:  Radiol Oncol       Date:  2016-02-16       Impact factor: 2.991

6.  Angular dependence correction of MatriXX and its application to composite dose verification.

Authors:  Yoshinobu Shimohigashi; Fujio Araki; Hirofumi Tominaga; Junichi Sakata; Keiichi Kawasaki; Nagisa Kanetake; Yuki Iwashita; Saori Yoshimura; Tamami Kawakami; Terunobu Ishihara; Tomoko Okuda; Kasei Kogo
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

7.  Evaluation of the efficiency and effectiveness of independent dose calculation followed by machine log file analysis against conventional measurement based IMRT QA.

Authors:  Baozhou Sun; Dharanipathy Rangaraj; Sunita Boddu; Murty Goddu; Deshan Yang; Geethpriya Palaniswaamy; Sridhar Yaddanapudi; Omar Wooten; Sasa Mutic
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

8.  Interplay effect of angular dependence and calibration field size of MapCHECK 2 on RapidArc quality assurance.

Authors:  Hosang Jin; Vance P Keeling; Daniel A Johnson; Salahuddin Ahmad
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

9.  Beam perturbation characteristics of a 2D transmission silicon diode array, Magic Plate.

Authors:  Ziyad A Alrowaili; Michael L F Lerch; Marco Petasecca; Martin G Carolan; Peter E Metcalfe; Anatoly B Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

10.  The MapCHECK Measurement Uncertainty function and its effect on planar dose pass rates.

Authors:  Daniel W Bailey; Jason D Spaans; Lalith K Kumaraswamy; Matthew B Podgorsak
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.