Abstract

INTRODUCTION: The cyclic fatigue of nickel-titanium (NiTi) rotary instruments has been studied extensively, but there is little information available on torsional fracture. Moreover, a clinical repeated locking effect was not considered in previous studies that evaluated torsional resistance of NiTi instruments. Thus, this study was aimed to compare the repetitive torsional resistance of various NiTi instruments with clinical relevance.
MATERIALS AND METHODS: Five brands of NiTi rotary instruments were selected: Twisted File (TF; SybronEndo, Orange, CA) and RaCe systems (FKG Dentaire, La Chaux-de-Fonds, Switzerland), both with an equilateral triangular cross-section, and the ProTaper (Dentsply Maillefer, Ballaignes, Switzerland), Helix (DiaDent, Chongju, Korea), and FlexMaster (VDW, Munchen, Germany), which had a convex triangular cross-section. Five millimeters of the tip of each file was embedded in composite resin block, and uniform torsional stresses (300 rpm, 1.0 N.cm) were applied repetitively by an endodontic motor with auto-stop mode until the file succumbed to torsional failure. The number of load applications leading to fracture was recorded. All fracture surfaces were examined under the SEM. Results were analyzed nonparametrically with alpha = 0.05.
RESULTS: Under the mode of load applications in this study, TF had the lowest and FlexMaster the highest torsional resistance among the groups (p < 0.05). Scanning electron microscopy examination revealed a typical pattern of torsional fracture for TF, RaCe, and ProTaper that was characterized by circular abrasion marks and skewed dimples near the center of rotation. In addition to these marks, Helix and FlexMaster presented a rough, torn-off appearance.
CONCLUSION: It was concluded that files of same cross-sectional design may exhibit different resistance to fracture probably as a result of the manufacturing process. Copyright 2010 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20630299     DOI: 10.1016/j.joen.2010.02.016

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  22 in total

1.  Assessment of the Centralization of Root Canal Preparation with Rotary Systems.

Authors:  Daniela Siqueira Lopes; Mariana Albuquerque Veiga Pessoa; Carlos Menezes Aguiar
Journal:  Acta Stomatol Croat       Date:  2016-09

2.  Canal transportation and centering ratio after preparation in severely curved canals: analysis by micro-computed tomography and double-digital radiography.

Authors:  Ane Poly; Fouad AlMalki; Fernando Marques; Bekir Karabucak
Journal:  Clin Oral Investig       Date:  2019-03-05       Impact factor: 3.573

3.  Cyclic fatigue resistance of four nickel-titanium rotary instruments: a comparative study.

Authors:  Eugenio Pedullà; Gianluca Plotino; Nicola Maria Grande; Alfio Pappalardo; Ernesto Rapisarda
Journal:  Ann Stomatol (Roma)       Date:  2012-08-09

4.  Effect of surface treatment on the mechanical properties of nickel-titanium files with a similar cross-section.

Authors:  Sang Won Kwak; Joo Yeong Lee; Hye-Jin Goo; Hyeon-Cheol Kim
Journal:  Restor Dent Endod       Date:  2017-06-28

5.  The Effect of Hydrofluoric Acid Surface Treatment on the Cyclic Fatigue Resistance of K3 NiTi Instruments.

Authors:  Kee-Yeon Kum; Seok Woo Chang
Journal:  Bioinorg Chem Appl       Date:  2017-05-02       Impact factor: 7.778

Review 6.  Rotary science and its impact on instrument separation: A focused review.

Authors:  Sandhya Anand Khasnis; Prem Prakash Kar; Apoorva Kamal; Jayaprakash D Patil
Journal:  J Conserv Dent       Date:  2018 Mar-Apr

7.  Cyclic fatigue resistance tests of Nickel-Titanium rotary files using simulated canal and weight loading conditions.

Authors:  Ok-In Cho; Antheunis Versluis; Gary Sp Cheung; Jung-Hong Ha; Bock Hur; Hyeon-Cheol Kim
Journal:  Restor Dent Endod       Date:  2013-02-26

8.  Comparison of cleaning efficiency and deformation characteristics of Twisted File and ProTaper rotary instruments.

Authors:  Hang Li; Chenzheng Zhang; Qing Li; Changning Wang; Yaling Song
Journal:  Eur J Dent       Date:  2014-04

9.  Resistance to torsional failure and cyclic fatigue resistance of ProTaper Next, WaveOne, and Mtwo files in continuous and reciprocating motion: An in vitro study.

Authors:  N O Varghese; Rajesh Pillai; U-Nu Sujathen; Shan Sainudeen; Abe Antony; Sinju Paul
Journal:  J Conserv Dent       Date:  2016 May-Jun

10.  Cyclic fatigue resistance, torsional resistance, and metallurgical characteristics of M3 Rotary and M3 Pro Gold NiTi files.

Authors:  Eugenio Pedullà; Fabio Lo Savio; Giusy Rita Maria La Rosa; Gabriele Miccoli; Elena Bruno; Silvia Rapisarda; Seok Woo Chang; Ernesto Rapisarda; Guido La Rosa; Gianluca Gambarini; Luca Testarelli
Journal:  Restor Dent Endod       Date:  2018-04-23
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.