Cyrus Zamani1, Xavi Illa, Sara Abdollahzadeh-Ghom, J R Morante, Albert Romano Rodríguez.
Abstract
Mesoporous silica with KIT-6 structure was investigated as a preconcentrating material in chromatographic systems for ammonia and trimethylamine. Its adsorption capacity was compared to that of existing commercial materials, showing its increased adsorption power. In addition, KIT-6 mesoporous silica efficiently adsorbs both gases, while none of the employed commercial adsorbents did. This means that KIT-6 Mesoporous silica may be a good choice for integrated chromatography/gas sensing micro-devices.Entities:
Keywords: Ammonia; Desorption; KIT-6; Mesoporous silica; Preconcentration; Trimethylamine
Year: 2009 PMID: 20628459 PMCID: PMC2894125 DOI: 10.1007/s11671-009-9396-5
Source DB: PubMed Journal: Nanoscale Res Lett ISSN: 1556-276X Impact factor: 4.703
Figure 1Schematic of the adsorbing assembly
Preconcentration settings
| Parameter | Set value |
|---|---|
| Sampling time | 60 s |
| Sampling temperature | 30 |
| Desorption temperature | 25 °C (for TMA), 100 °C (for NH3) |
| Desorption rate | Maximum (automatically preformed by preconcentrating device) |
Figure 2Chromatogram ofasynthetic air, andbTMA (presenting peaks of air and TMA) analyzed in “Plot A”. In chromatogramapeaks of air and humidity were observed. Passing TMA through preconcentrator (Chromatogramb) results in TMA adsorption/desorption with a large peak
Chromatograph settings
| “Plot A” | “Plot U” | |
|---|---|---|
| Carrier gas | He | He |
| Column temperature | 100 °C | 100 °C |
| Sampling time | 15 s | 15 s |
| Analysis time | Variable (Max. 600 s) | Variable (Max. 600 s) |
| Column pressure | 30 Psi | 25 Psi |
| Sensitivity | High | High |
Figure 3Comparison of TMA peaks obtained using KIT-6 and Carboxen 569 detected by “Plot A”. Both KIT-6 and commercial material show the same peak height
Figure 4Comparison of KIT-6, Hayesep P, and Carboxen 569 for NH3preconcentration and “Plot A”. Compared to both Carboxen and Hayesep, KIT-6 shows a much larger signal
Figure 5Higher response of KIT-6 when compared to refor TMA detection in “relot A”. In addition to TMA, peaks of air, CO2,and humidity are also seen. The small peak received just before TMA belongs to the small amount of TMA in the dead volume
NH3peaks preconcentrated by KIT-6 and Hayesep P detected by Micro GC
| “Plot A” | “Plot U” | ||
|---|---|---|---|
| KIT-6 | Height | 39,000 | 9,000 |
| Area | 24,500 | 18,500 | |
| Hayesep P | Height | 18,000 | 2,000 |
| Area | 9,000 | 4,000 | |