Literature DB >> 20616404

A two-dose-rate method for general recombination correction for liquid ionization chambers in pulsed beams.

Heikki Tölli1, Rickard Sjögren, Mikael Wendelsten.   

Abstract

The correction for general recombination losses in liquid ionization chambers (LICs) is more complex than that in air-filled ionization chambers. The reason for this is that the saturation charge in LICs, i.e. the charge that escapes initial recombination, depends on the applied voltage. This paper presents a method, based on measurements at two different dose rates in a pulsed beam, for general recombination correction in LICs. The Boag theory for pulsed beams is used and the collection efficiency is determined by numerical methods which are equivalent to the two-voltage method used in dosimetry with air-filled ionization chambers. The method has been tested in experiments in water in a 20 MeV electron beam using two LICs filled with isooctane and tetramethylsilane. The dose per pulse in the electron beam was varied between 0.1 mGy/pulse and 8 mGy/pulse. The relative standard deviations of the collection efficiencies determined with the two-dose-rate method ranged between 0.1% and 1.5%. The dose-rate variations of the general recombination corrected charge measured with the LICs are in excellent agreement with the corresponding values obtained with an air-filled plane parallel ionization chamber.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20616404     DOI: 10.1088/0031-9155/55/15/004

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  General ion recombination effect in a liquid ionization chamber in high-dose-rate pulsed photon and electron beams.

Authors:  Hideyuki Takei; Tomonori Isobe; Nozomi Kitamura; Yutaro Mori; Tetsuya Tomita; Daisuke Kobayashi; Satoshi Kamizawa; Tomoharu Sato; Hideyuki Sakurai; Takeji Sakae
Journal:  J Radiat Res       Date:  2018-05-01       Impact factor: 2.724

2.  Pretreatment Dose Verification in Volumetric Modulated Arc Therapy Using Liquid Ionization Chamber.

Authors:  Nitin R Kakade; Rajesh Kumar; Sunil Dutt Sharma; Vikram Mittal; D Datta
Journal:  J Med Phys       Date:  2019 Jan-Mar

3.  Characterization of recombination effects in a liquid ionization chamber used for the dosimetry of a radiosurgical accelerator.

Authors:  Antoine Wagner; Frederik Crop; Thomas Lacornerie; Nick Reynaert
Journal:  J Vis Exp       Date:  2014-05-09       Impact factor: 1.355