Predictors of inappropriate atrial sensing in long-term VDD-pacing systems.

Chin-Chou Huang1, Ta-Chuan Tuan, Man-Cai Fong, Wen-Shin Lee, Chi-Woon Kong.   

Abstract

AIMS: The efficacy and stability of the atrial electrode sensing function is essential for maintaining atrioventricular (AV) synchrony. This study aimed to explore the long-term reliability and causes of the long-term sensing failure of VDD systems. METHODS AND
RESULTS: We enrolled all the patients with complete or high-degree AV block who received VDD pacemakers between August 1994 to January 2006 and who were followed up for more than 12 months. The interrogation parameters, including the atrial potentials (APs) and AV-synchrony ratio were acquired immediately post-implantation and at 3-6 month intervals thereafter. An inappropriate atrial sensing efficacy was defined as an AV-synchrony ratio of <90%. Totally 157 patients (70 +/- 12 years, 103 males) were enrolled into the study with a follow-up for 4.9 +/- 2.5 years. Twenty-six patients (16.6%) suffered from inappropriate atrial sensing. According to a Kaplan-Meier analysis, the incidence of inappropriate atrial sensing was higher in the patients with an age > or =72 years old (P = 0.047), mean AP during the implantation of <3.0 mV (P = 0.015), concomitant use of non-dihydropyridine calcium channel blockers (CCBs) (P = 0.003), and atrial fibrillation (AF) (P < 0.001). A Cox regression analysis showed that non-dihydropyridine CCBs (hazard ratio, 3.255; 95% confidence interval, 1.148-9.227, P = 0.026) and AF (hazard ratio, 6.507; 95% confidence interval, 2.478-17.104, P < 0.001) predicted inappropriate atrial sensing.
CONCLUSION: VDD pacing is a reliable pacing modality. However, we should monitor the pacemaker sensing function in the patients with the concomitant use of non-dihydropyridine CCBs and AF.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20601673     DOI: 10.1093/europace/euq190

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  2 in total

1.  Long-term outcome in patients receiving permanent pacemaker implantation for atrioventricular block: Comparison of VDD and DDD pacing.

Authors:  Jo-Nan Liao; Tze-Fan Chao; Ta-Chuan Tuan; Chi-Woon Kong; Shih-Ann Chen
Journal:  Medicine (Baltimore)       Date:  2016-08       Impact factor: 1.889

2.  Echocardiographic Parameters as Predictors for the Efficiency of Resynchronization Therapy in Patients with Dilated Cardiomyopathy and HFrEF.

Authors:  Silvius-Alexandru Pescariu; Raluca Şoşdean; Cristina Tudoran; Adina Ionac; Gheorghe Nicusor Pop; Romulus Zorin Timar; Sorin Pescariu; Mariana Tudoran
Journal:  Diagnostics (Basel)       Date:  2021-12-24
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.