Transarterial pacemaker lead implantation results in acute myocardial infarction.

Ziad F Issa1, John B Gill.   

Abstract

We report a case of inadvertent transarterial implantation of dual-chamber pacemaker leads; the ventricular lead positioned across the aortic valve into the left ventricle and the atrial lead curving in the aortic root with the tip positioned into the left circumflex artery, resulting in acute myocardial infarction. The diagnosis was made based on the finding on the chest X-ray, surface ECG, and coronary angiography.

Entities:  

Mesh:

Year:  2010        PMID: 20601362     DOI: 10.1093/europace/euq189

Source DB:  PubMed          Journal:  Europace        ISSN: 1099-5129            Impact factor:   5.214


  2 in total

Review 1.  Beware of the coronary arteries with implantable cardiac electronic devices.

Authors:  S Serge Barold; Benjamin J Pang; Harry G Mond
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2017-09

2.  Shift From Left to a Right Bundle Block on ECG Leading to the Diagnosis of a Malpositioned Lead in the Coronary Sinus: A Case Report.

Authors:  Pramod Theetha Kariyanna; Yuvraj Singh Chowdhury; Amog Jayarangaiah; Jonathan Christopher Francois; Pakinam Mekki; Isabel M McFarlane
Journal:  Am J Med Case Rep       Date:  2020-04-14