| Literature DB >> 20595890 |
Hannah Zimmerman1, Romana Coehlo-Anderson, Marvin Slepian, Richard G Smith, Gulshan Sethi, Jack G Copeland.
Abstract
We report two cases at a single institution, a 52-year-old man and a 25-year-old woman, who had failures of their CardioWest total artificial hearts (TAH) from central venous lines that caused obstruction of the right ventricular inflow valves. Each patient was noted to have decreasing TAH outputs for a few days before this catastrophic device arrest. Both patients died; one suddenly and one during a protracted period, as a result of catheter entrapment of the disc of the Medtronic Hall right ventricular inflow valve.Entities:
Mesh:
Year: 2010 PMID: 20595890 DOI: 10.1097/MAT.0b013e3181e87d4f
Source DB: PubMed Journal: ASAIO J ISSN: 1058-2916 Impact factor: 2.872