BACKGROUND: Calibration of a microarray scanner is critical for accurate interpretation of microarray results. Shi et al. (BMC Bioinformatics, 2005, 6, Art. No. S11 Suppl. 2.) reported usage of a Full Moon BioSystems slide for calibration. Inspired by the Shi et al. work, we have calibrated microarray scanners in our previous research. We were puzzled however, that most of the signal intensities from a biological sample fell below the sensitivity threshold level determined by the calibration slide. This conundrum led us to re-investigate the quality of calibration provided by the Full Moon BioSystems slide as well as the accuracy of the analysis performed by Shi et al.Entities:
Mesh:
Year: 2010 PMID: 20594322 PMCID: PMC2908103 DOI: 10.1186/1471-2105-11-361
Source DB: PubMed Journal: BMC Bioinformatics ISSN: 1471-2105 Impact factor: 3.169
Calibration slide description provided by Full Moon BioSystems.
| Series | Fluorophores/um2 | Series | Fluorophores/um2 |
|---|---|---|---|
| 1 | 1.47E + 05 | 17 | 2.24E+00 |
| 2 | 7.35E + 04 | 18 | 1.12E+00 |
| 3 | 3.68E + 04 | 19 | 5.61E-01 |
| 4 | 1.84E + 04 | 20 | 2.80E-01 |
| 5 | 9.19E + 03 | 21 | 1.40E-01 |
| 6 | 4.59E + 03 | 22 | 7.01E-02 |
| 7 | 2.30E + 03 | 23 | 3.50E-02 |
| 8 | 1.15E + 03 | 24 | 1.75E-02 |
| 9 | 5.74E + 02 | 25 | 8.76E-03 |
| 10 | 2.87E + 02 | 26 | 4.38E-03 |
| 11 | 1.44E + 02 | 27 | 2.19E-03 |
| 12 | 7.18E + 01 | ||
| 13 | 3.59E + 01 | 29 | 0 |
| 14 | 1.79E + 01 | 30 | 0 |
| 15 | 8.97E + 00 | 31 | 0 |
| 16 | 4.49E + 00 | 32 | Position Marker |
Layout file downloaded from the Full Moon BioSystems web site.
| Column | Name | ID |
|---|---|---|
| 1 | Marker | Cy3 |
| 2 | 0 | Cy3 |
| 3 | 0 | Cy3 |
| 4 | 0 | Cy3 |
| Cy3 | ||
| 6 | 2.19E-03 | Cy3 |
| 7 | 4.38E-03 | Cy3 |
| 8 | 8.76E-03 | Cy3 |
| ... | ... | ... |
Figure 1Photomultiplier response of the Agilent scanner, Cy3 channel. Photomultiplier gain is 100.
Characteristics of microarray scanners.
| Scanner | Channel | PMTi | gii | ||
|---|---|---|---|---|---|
| Agilent | Cy3 | 100 | 106 | 0.95 | 11.90 |
| Agilent | Cy3 | 50 | 52 | 0.95 | 12.14 |
| Agilent | Cy3 | 20 | 21 | 0.94 | 12.12 |
| Agilent | Cy5 | 100 | 54 | 0.98 | 3.90 |
| Agilent | Cy5 | 50 | 28 | 0.98 | 4.04 |
| Agilent | Cy5 | 20 | 11 | 0.99 | 4.76 |
| GenePix | Cy3 | 500 | 208 | 0.74 | 7.45 |
| Bio-Rad | Cy3 | 500 | 4 | 0.84 | 8.33 |
| Bio-Rad | Cy5 | 500 | 11 | 0.86 | 3.85 |
i PMT settings are particular for each instrument and should not be compared between the instruments.
ii The dimension of the g value is in Fluorophores/um2.
Figure 2Comparison of PMT responses for different channels and gains for Agilent (top row) and Bio-Rad (bottom row) scanners. Left column shows data in log-log coordinates, right column - in original non-transformed coordinates.