Ji Hyun Chin1, Eun Ho Lee, Dae Kee Choi, In Cheol Choi.
Abstract
Airway management during carinal resection should provide adequate ventilation and oxygenation as well as a good surgical field, but without complications such as barotraumas or aspiration. One method of airway management is high frequency jet ventilation (HFJV) of one lung or both lungs. We describe a patient undergoing carinal resection, who was managed with HFJV of one lung, using a de-ballooned bronchial blocker of a Univent tube without cardiopulmonary compromise. HFJV of one lung using a bronchial blocker of a Univent tube is a simple and safe method which does not need additional catheters to perform HFJV and enables the position of the stiffer bronchial blocker more stable in airway when employed during carinal resection.Entities:
Keywords: Carinal Resection; High Frequency Jet Ventilation; One Lung Ventilation; Univent Tube