| Literature DB >> 20590757 |
Shomei Ryozawa1, Hirotoshi Iwano, Kumiko Taba, Manabu Senyo, Isao Sakaida.
Abstract
We report the successful retrieval of an impacted mechanical lithotripsy basket. In a patient with two large common bile duct stones, the basket with the entrapped stone was impacted within the mid-common bile duct. We then attempted to use another mechanical lithotripter; however the central wire of the basket fractured at the handle portion. Grasping a few wires of the impacted basket with rat-tooth forceps allowed the wires of the basket to slip away from the stone. The present report describes the safe and effective use of rat-tooth forceps in the management of an impacted lithotripter basket.Entities:
Mesh:
Year: 2010 PMID: 20590757 DOI: 10.1111/j.1443-1661.2010.00963.x
Source DB: PubMed Journal: Dig Endosc ISSN: 0915-5635 Impact factor: 7.559