Patient state index vs bispectral index as measures of the electroencephalographic effects of propofol.

M Soehle1, M Kuech, M Grube, S Wirz, S Kreuer, A Hoeft, J Bruhn, R K Ellerkmann.   

Abstract

BACKGROUND: The patient state index (PSI) and the bispectral index (BIS) quantify anaesthetic depth based on the EEG using different algorithms. We compared both indices with regard to the prediction of the depth of propofol anaesthesia.
METHODS: In 17 patients, propofol was infused until burst suppression occurred and stopped thereafter until BIS recovered to values above 60. This was repeated; afterwards, patients were intubated, for subsequent surgery. Without surgical stimulus, PSI and BIS were measured simultaneously and compared with the estimated effect-site concentrations of propofol. These were derived from simultaneous pharmacokinetic and -dynamic modelling in an individual two-stage and a population-based NONMEM approach.
RESULTS: A close sigmoid relationship was observed between the propofol effect-site concentration and both PSI [coefficient of determination rho(2)=0.91 (sd 0.05)] and BIS [rho(2)=0.92 (0.03)], which was significantly steeper for PSI [gamma=2.2 (0.6)] than for BIS [gamma=1.8 (0.4)], and reached significantly lower values for PSI [E(max)=0.3 (1.1)] than for BIS [E(max)=5.3 (6.7)] at maximal propofol concentrations. A significantly smaller k(e0) was obtained for PSI [0.09 (0.03) min(-1)] compared with BIS [0.10 (0.02) min(-1)]. PSI and BIS correlated significantly with each other (rho(2)=0.866) and predicted propofol effect-site concentration with a comparable probability [P(K)=0.87 (0.05) and 0.86 (0.05), respectively]. NONMEM revealed E(0)=89.3 and 92.3, E(max)=1.9 and 8.6, C(e50)=1.38 and 1.92 microg ml(-1), gamma=1.6 and 1.48, and k(e0)=0.103 and 0.131 min(-1) as typical values for PSI and BIS, respectively.
CONCLUSIONS: The PSI and the BIS monitors performed equally well in predicting depth of propofol anaesthesia. However, PSI was lower than BIS by approximately 10-15 points at high propofol concentrations.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20587537     DOI: 10.1093/bja/aeq155

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  12 in total

1.  Oxygen reserve index (ORi™) contributes to prediction of hypoxemia and patient safety during tracheal stent insertion using rigid bronchoscopy: a case report.

Authors:  Yasunori Niwa; Juntaro Shiba; Hiromasa Fujita; Risa Oka; Mamoru Takeuchi
Journal:  J Clin Monit Comput       Date:  2018-12-05       Impact factor: 2.502

2.  The Patient State Index is well balanced for propofol sedation.

Authors:  K H Lee; Y H Kim; Y J Sung; M K Oh
Journal:  Hippokratia       Date:  2015 Jul-Sep       Impact factor: 0.471

3.  Electroencephalogram-Based Evaluation of Impaired Sedation in Patients with Moderate to Severe COVID-19 ARDS.

Authors:  Armin Niklas Flinspach; Sebastian Zinn; Kai Zacharowski; Ümniye Balaban; Eva Herrmann; Elisabeth Hannah Adam
Journal:  J Clin Med       Date:  2022-06-17       Impact factor: 4.964

4.  Burst-suppression ratio underestimates absolute duration of electroencephalogram suppression compared with visual analysis of intraoperative electroencephalogram.

Authors:  W G Muhlhofer; R Zak; T Kamal; B Rizvi; L P Sands; M Yuan; X Zhang; J M Leung
Journal:  Br J Anaesth       Date:  2017-05-01       Impact factor: 9.166

5.  Intraoperative burst suppression is associated with postoperative delirium following cardiac surgery: a prospective, observational study.

Authors:  Martin Soehle; Alexander Dittmann; Richard K Ellerkmann; Georg Baumgarten; Christian Putensen; Ulf Guenther
Journal:  BMC Anesthesiol       Date:  2015-04-28       Impact factor: 2.217

6.  Comparison of Two Methods of Anesthesia Using Patient State Index: Propofol Versus Sevoflurane During Interventional Neuroradiology Procedure.

Authors:  Eunsu Kang; Ki Hwa Lee; Jae-Hong Park
Journal:  Anesth Pain Med       Date:  2019-04-16

7.  Study of the rational dose of propofol in elderly patients under bispectral index monitoring during total intravenous anesthesia: A PRISMA-compliant systematic review.

Authors:  Lina Jia; Jiachen Hou; Haibo Zheng; Lihua Sun; Yingying Fan; Xu Wang; Mingyue Hao; Yue Li; Tongwei Yang
Journal:  Medicine (Baltimore)       Date:  2020-01       Impact factor: 1.889

8.  Simultaneous comparison of depth of sedation performance between SedLine and BIS during general anesthesia using custom passive interface hardware: study protocol for a prospective, non-blinded, non-randomized trial.

Authors:  James Harvey Jones; Vinay Ravikumar Nittur; Neal Fleming; Richard L Applegate
Journal:  BMC Anesthesiol       Date:  2021-04-06       Impact factor: 2.217

9.  Processed Multiparameter Electroencephalogram-Guided General Anesthesia Management Can Reduce Postoperative Delirium Following Carotid Endarterectomy: A Randomized Clinical Trial.

Authors:  Na Xu; Li-Xia Li; Tian-Long Wang; Li-Qun Jiao; Yang Hua; Dong-Xu Yao; Jie Wu; Yan-Hui Ma; Tian Tian; Xue-Li Sun
Journal:  Front Neurol       Date:  2021-07-12       Impact factor: 4.003

10.  Comparison of Bispectral Index and Patient State Index values according to recovery from moderate neuromuscular block under steady-state total intravenous anesthesia.

Authors:  Doyeon Kim; Jin Hee Ahn; Gunyoung Heo; Ji Seon Jeong
Journal:  Sci Rep       Date:  2021-03-15       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.