Reliability and validity of a new ankle electrogoniometer.

S Bronner1, S Agraharasamakulam, S Ojofeitimi.   

Abstract

OBJECTIVES: To establish reliability, accuracy and concurrent validity of a new electrogoniometer (SG110A, Biometrics) for measuring ankle movement.
METHODS: The new electrogoniometer, placed at the lateral malleolus (A-perp), was compared to a customary electrogoniometer (SG110, Biometrics), placed along the Achilles tendon (A-para). Concurrent recordings were made with motion analysis. Common dance movements were performed by 17 dancers on two days. Intraclass correlation coefficients (ICC) and standard error of measurement (SEM) were calculated for sagittal angular displacements.
RESULTS: Instrument and intra-tester reliability of A-perp was high (r >or= 0.985, r >or= 0.979, respectively). Criterion and concurrent validity were also high (r >or= 0.954, r >or= 0.937). SEM ranged from 1.43 degrees to 6.99 degrees.
CONCLUSIONS: The high ICC values establish acceptable reliability and validity. Subjects found A-perp more comfortable. Although both electrogoniometers are acceptable (comparable SEM) to measure extreme ankle motions, the greater comfort and durability of the new A-perp sensor is advantageous.

Mesh:

Year:  2010        PMID: 20586555     DOI: 10.3109/03091902.2010.493968

Source DB:  PubMed          Journal:  J Med Eng Technol        ISSN: 0309-1902


  5 in total

1.  Does the type of visual feedback information change the control of standing balance?

Authors:  Fabio Dos Anjos; Thiago Lemos; Luís Aureliano Imbiriba
Journal:  Eur J Appl Physiol       Date:  2016-07-18       Impact factor: 3.078

2.  Assessment of Correlation between Electrogoniometer Measurements and Sports-Specific Movement in Karate Elites.

Authors:  Sara Lotfian; Afsaneh Safar Cherati; Aliashraf Jamshidi; Mohammad Ali Sanjari
Journal:  Asian J Sports Med       Date:  2014-06

3.  Ankle Kinematics and Temporal Gait Characteristics over the Duration of a 6-Minute Walk Test in People with Multiple Sclerosis Who Experience Foot Drop.

Authors:  Marietta L van der Linden; Georgia Andreopoulou; Judy Scopes; Julie E Hooper; Thomas H Mercer
Journal:  Rehabil Res Pract       Date:  2018-07-02

4.  Validity and Reliability of an Inertial Device for Measuring Dynamic Weight-Bearing Ankle Dorsiflexion.

Authors:  José M Oliva-Lozano; Isabel Martín-Fuentes; José M Muyor
Journal:  Sensors (Basel)       Date:  2020-01-10       Impact factor: 3.576

5.  Sensory Stimulation of the Foot and Ankle Early Post-stroke: A Pilot and Feasibility Study.

Authors:  Alison M Aries; Valerie M Pomeroy; Julius Sim; Susan Read; Susan M Hunter
Journal:  Front Neurol       Date:  2021-07-05       Impact factor: 4.003

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.