Intravenous fluid resuscitation: was Poiseuille right?

Andrew D Reddick1, Julie Ronald, William G Morrison.   

Abstract

AIM: To compare the flow rates of readily available intravenous infusion devices and to compare the effect of the addition of pressure or a needle-free intravenous connector device.
METHODS: Several intravenous devices with different characteristics had their flow rates determined under a standard set of conditions. The flow rates were then measured with the addition of a pressure bag to the system and then with a needle-free intravenous connector device. The flow rates and change in flow rates were then analysed.
RESULTS: The results showed a general agreement with Poiseuille's law. The needle-free connector slowed the rate of flow by up to 41.4% with the greatest effect on short, wide-bore devices. The addition of pressure had a greater effect on longer devices.
CONCLUSIONS: Short, wide cannulae should be used when rapid fluid resuscitation is required. Needle-free devices should not be used when rapid fluid resuscitation is needed.

Mesh:

Year:  2010        PMID: 20581377     DOI: 10.1136/emj.2009.083485

Source DB:  PubMed          Journal:  Emerg Med J        ISSN: 1472-0205            Impact factor:   2.740


  5 in total

1.  Basic Physics Revisited for a Surgeon.

Authors:  J S Rajkumar; Pradeep Chopra
Journal:  Indian J Surg       Date:  2015-06       Impact factor: 0.656

2.  Paediatric cerebrovascular CT angiography-towards better image quality.

Authors:  Stefanie C Thust; Wui Khean Kling Chong; Roxana Gunny; Asif Mazumder; Marius Poitelea; Anna Welsh; Ash Ederies; Kshitij Mankad
Journal:  Quant Imaging Med Surg       Date:  2014-12

3.  Rapid fluid administration: an evaluation of two techniques.

Authors:  Holly C Gillis; Hina Walia; Dmitry Tumin; Tarun Bhalla; Joseph D Tobias
Journal:  Med Devices (Auckl)       Date:  2018-09-17

4.  Beyond poiseuille: preservation fluid flow in an experimental model.

Authors:  Saurabh Singh; Lucy V Randle; Paul T Callaghan; Christopher J E Watson; Chris J Callaghan
Journal:  J Transplant       Date:  2013-08-26

5.  Pressure-related flow rates for continuous renal replacement therapy in very small children: an in vitro study.

Authors:  Isabella Stevens-Harris; Dusan Raffaj; Patrick Davies
Journal:  BMJ Paediatr Open       Date:  2017-08-04
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.