Literature DB >> 20574760

[Cleft in carbon dioxide absorber. Intraoperative problems with ventilation due to a leak in the breathing circuit].

C Paul1, B W Böttiger.   

Abstract

In the case presented problems with mechanical and manual ventilation of a patient occurred during the operation. Prior to this endotracheal intubation had been performed without difficulty and the respirator had passed all system checks. A leakage in the recently changed carbon dioxide absorber could be detected which had been accidentally dropped and damaged internally.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20574760     DOI: 10.1007/s00101-010-1751-5

Source DB:  PubMed          Journal:  Anaesthesist        ISSN: 0003-2417            Impact factor:   1.041


  5 in total

1.  [Post-lobectomy bronchopleural fistula -- a challenge for postoperative intensive care].

Authors:  T Mendel; J Jakubetz; M Steen; R Stuttmann
Journal:  Anasthesiol Intensivmed Notfallmed Schmerzther       Date:  2006-04       Impact factor: 0.698

2.  [Accidentally opened adjustable pressure-limiting valve. Failure of manual ventilation].

Authors:  I Eggemann; B W Böttiger; F Spöhr
Journal:  Anaesthesist       Date:  2009-03       Impact factor: 1.041

3.  Leak in the breathing circuit: CO2 absorber and human error.

Authors:  Goneppanavar Umesh; Kaur Jasvinder; Roy Sagarnil
Journal:  J Clin Monit Comput       Date:  2010-02-13       Impact factor: 2.502

4.  Defective carbon dioxide absorber as a cause for a leak in a breathing circuit.

Authors:  A M Kshatri; C P Kingsley
Journal:  Anesthesiology       Date:  1996-02       Impact factor: 7.892

Review 5.  How a system for reporting medical errors can and cannot improve patient safety.

Authors:  John R Clarke
Journal:  Am Surg       Date:  2006-11       Impact factor: 0.688

  5 in total
  2 in total

1.  Failure to replace a CO2-absorber due to shifted metal pins: a potential source of hazard to patients.

Authors:  Hendrik Drinhaus; Valentina Rahn; Thorsten Annecke
Journal:  J Clin Monit Comput       Date:  2017-06-16       Impact factor: 2.502

Review 2.  Anaesthesia machine: checklist, hazards, scavenging.

Authors:  Umesh Goneppanavar; Manjunath Prabhu
Journal:  Indian J Anaesth       Date:  2013-09