Literature DB >> 20563163

Longterm changes of silicon photodiodes and their use for photometric standardization.

G Eppeldauer.   

Abstract

A secondary standard silicon photodiode matched with a V-lambda filter was calibrated against primary standard, self-calibrated inversion layer silicon photodiodes, to achieve a high accuracy photometer, according to the new definition of the canadela (the photometric base unit). The measured several percent/year specular spectral reflectance change of the windowless primary standard photodiodes was eliminated by their repeated self-calibration. This self-calibration also eliminated the measured several tenth of a percent/year spectral response change of the secondary standard silicon photodiode. The secondary standard detector could be a nonunity quantum efficiency light detector. The spectral response calibration of the V-lambda matched detector of medium spectral mismatch (f(1)= 3.0%) against the absolute spectral responses of three self-calibrated photodiodes resulted in a standard deviation of 0.17% in luminous flux (lumen) calibration. Also illuminance (lux) and light intensity (candela) calibrations were derived from the above primary photometric calibration. It is shown that the V-lambda matched photometer with the above spectral calibration can be used for accurate photometric measurements for all kinds of light sources of known spectral power distribution.

Entities:  

Year:  1990        PMID: 20563163     DOI: 10.1364/AO.29.002289

Source DB:  PubMed          Journal:  Appl Opt        ISSN: 1559-128X            Impact factor:   1.980


  1 in total

1.  The NIST Detector-Based Luminous Intensity Scale.

Authors:  C L Cromer; G Eppeldauer; J E Hardis; T C Larason; Y Ohno; A C Parr
Journal:  J Res Natl Inst Stand Technol       Date:  1996 Mar-Apr
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.