Literature DB >> 20558847

Risk of hematoma complications after device implant in the clopidogrel era.

Ilana B Kutinsky1, Regina Jarandilla, Maralee Jewett, David E Haines.   

Abstract

BACKGROUND: Device implant pocket hematoma is a recognized complication after permanent pacemaker (PM) and implantable cardioverter-defibrillator (ICD) implantation. Pocket hematoma is associated with local discomfort, an increased risk of infection, and may require surgical intervention or lead to lengthier hospital stays. The purpose of the study was to identify the clinical factors associated with hematoma formation after PM or ICD device implantation. METHODS AND
RESULTS: The subjects of this prospective observational study were 935 consecutive patients at Beaumont Hospital who underwent implantation of a PM or an ICD. Clinical characteristics and anticoagulant/antiplatelet drug use were recorded. A pocket hematoma was documented in 89 of 935 patients. Significant predictors of device pocket hematoma included ongoing clopidogrel therapy (18.3% on therapy, 10.5% recently discontinued, and 7.9% off therapy; P<0.001) and use of intravenous heparin (22.0% on therapy versus 8.2%; P<0.0001). Patients in whom clopidogrel was discontinued >4 days before device implantation had no hematoma. Hematomas occur more frequently in patients receiving ICDs than those receiving PMs. Device pocket hematoma was associated with an increased median length of hospital stay (4 days [interquartile range, 1 to 9] days with versus 2 days [ interquartile range, 1 to 6] days without hematoma; P=0.004) and increased late complications or surgical intervention.
CONCLUSIONS: The use of clopidogrel or intravenous heparin significantly increased the risk of hematoma at the time of PM or ICD implantation. By withholding clopidogrel before surgery, the excess risk of bleeding complications may be reduced.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20558847     DOI: 10.1161/CIRCEP.109.917625

Source DB:  PubMed          Journal:  Circ Arrhythm Electrophysiol        ISSN: 1941-3084


  19 in total

Review 1.  Meta-analysis of bleeding complications associated with cardiac rhythm device implantation.

Authors:  Michael L Bernard; Matthew Shotwell; Paul J Nietert; Michael R Gold
Journal:  Circ Arrhythm Electrophysiol       Date:  2012-04-24

Review 2.  Ablation of Longstanding Persistent Atrial Fibrillation.

Authors:  Edward J Ciaccio
Journal:  J Atr Fibrillation       Date:  2010-10-22

3.  Implantable electrophysiologic cardiac device infections: a risk factor analysis.

Authors:  D Raad; J Irani; E G Akl; S Choueiri; E Azar; J Abboud; C Afif
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2012-08-25       Impact factor: 3.267

Review 4.  Perioperative management of antithrombotic therapy in patients receiving cardiovascular implantable electronic devices: a network meta-analysis.

Authors:  Hua He; Bing-Bing Ke; Yan Li; Fu-Sheng Han; Xiaodong Li; Yu-Jie Zeng
Journal:  J Interv Card Electrophysiol       Date:  2017-08-25       Impact factor: 1.900

Review 5.  Arguments to Apply Epinephrine for Pocket Hematoma Reduction. The MAITRE Study.

Authors:  Nikolay Ilov; Nikolay Ilov; Anatoly Nechepurenko; Albert Abdulkadyrov; Damir Paskeev; Elena Damrina; Elena Kulikova; Marina Terent'eva; Dinara Stompel; Dmitry Tarasov
Journal:  J Atr Fibrillation       Date:  2016-06-30

6.  Use of a pocket compression device for the prevention and treatment of pocket hematoma after pacemaker and defibrillator implantation (STOP-HEMATOMA-I).

Authors:  Mohit K Turagam; Darbhamulla V Nagarajan; Krzysztof Bartus; Akash Makkar; Vijay Swarup
Journal:  J Interv Card Electrophysiol       Date:  2017-04-17       Impact factor: 1.900

7.  Influence of oral antiplatelet therapy on hemorrhagic complications of pacemaker implantation.

Authors:  Samir M Said; Hans D Esperer; Judit Hahn; Andreas Bollmann; Sergio Richter; Thomas Rauwolf; Roland Prondzinsky; Alexander Schmeisser; Ruediger C Braun-Dullaeus
Journal:  Clin Res Cardiol       Date:  2013-02-06       Impact factor: 5.460

8.  Effectiveness of a Non-Taped Compression Dress in Patients Receiving Cardiac Implantable Electronic Devices.

Authors:  Chih-Yin Chien; Ya-Hui Chang; Yih-Jer Wu; Po-Lin Lin; Wei-Ru Chiou; Bo-Ching Chi; Hsu-Ping Wu; Feng-Ching Liao; Ying-Hsiang Lee
Journal:  Acta Cardiol Sin       Date:  2019-05       Impact factor: 2.672

9.  Prevention of cardiac implantable electronic device infections: guidelines and conventional prophylaxis.

Authors:  Carina Blomstrom-Lundqvist; Bozena Ostrowska
Journal:  Europace       Date:  2021-05-25       Impact factor: 5.214

10.  Postimplantation pocket hematoma increases risk of cardiac implantable electronic device infection: A meta-analysis.

Authors:  Jakrin Kewcharoen; Chanavuth Kanitsoraphan; Sittinun Thangjui; Thiratest Leesutipornchai; Sakditad Saowapa; Apichai Pokawattana; Leenhapong Navaravong
Journal:  J Arrhythm       Date:  2021-03-13