Microgap between zirconia abutments and titanium implants.

Sebastien Baixe1, Gabriel Fauxpoint, Youri Arntz, Olivier Etienne.   

Abstract

PURPOSE: The aim of this study was to evaluate in vitro the microgap between different zirconia abutments and their titanium implants.
MATERIALS AND METHODS: Four systems were evaluated: Procera zirconia (Nobel Biocare) (Nb), Cercon Balance Anterior (Dentsply Friadent) (Ba), ZirDesign (Astratech) (Zd), and Straumann Cares ceramic (Straumann) (Ca). Five assemblies were assessed for each system. The assemblies were embedded in epoxy, cut along their long axes, and polished. Scanning electron microscopic observations were made along the first 100 microm of the gap on each side at maximal magnification. Images were combined and gap measurements were made 10 microm apart. A two-way analysis of variance was performed on the data.
RESULTS: Scanning electron micrographs showed a mean marginal microgap of 0.89 microm (SD 1.67) for all assemblies. Significant differences (P < .001) were observed between mean (+/- SD) microgap measurements of the four tested systems: Ba = 0.38 +/- 0.28 microm; Zd = 0.55 +/- 0.23 microm; Nb = 1.83 +/- 3.21 microm; Ca = 0.90 +/- 0.59 microm. The mean microgap of the first 20 microm of the outer region (1.66 microm) was significantly (P < .001) larger than the mean microgap (0.56 microm) of the inner region (30 to 100 microm).
CONCLUSIONS: Within the limitations of this study, the mean microgap observed for all tested systems was less than 2 microm. For each system, the microgap decreased quickly from the outer region to the inner. The mean gap was larger for flat-to-flat connection systems, compared to internal-connection systems with a conical interface. These results demonstrate smaller microgaps compared to those described in the literature for titanium abutments. The precise fit of these abutments could lead to better biologic and biomechanical behavior.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20556243

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  4 in total

1.  Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and external-hex implant connections.

Authors:  Hakimeh Siadat; Elaheh Beyabanaki; Niloufar Mousavi; Marzieh Alikhasi
Journal:  J Adv Prosthodont       Date:  2017-08-16       Impact factor: 1.904

2.  Sphere-plane methodology to evaluate the wear of titanium of dental implants: a research proposal.

Authors:  Teresa Almeida Mendes; João Caramês; Luís Pires Lopes; Amílcar Lopes Ramalho
Journal:  BMC Res Notes       Date:  2018-07-31

3.  Influence of Implant Connection, Abutment Design and Screw Insertion Torque on Implant-Abutment Misfit.

Authors:  Jorge Vélez; Jesús Peláez; Carlos López-Suárez; Rubén Agustín-Panadero; Celia Tobar; María J Suárez
Journal:  J Clin Med       Date:  2020-07-24       Impact factor: 4.241

4.  Implant-Abutment Misfit after Cyclic Loading: An In Vitro Experimental Study.

Authors:  John Eversong Lucena de Vasconcelos; Jefferson David Melo de Matos; Daher Antonio Queiroz; Guilherme da Rocha Scalzer Lopes; Bruna Caroline Gonçalves Vasconcelos de Lacerda; Marco Antonio Bottino; Cecilia Pedroso Turssi; Roberta Tarkany Basting; Flávia Lucisano Botelho do Amaral; Fabiana Mantovani Gomes França
Journal:  Materials (Basel)       Date:  2022-08-03       Impact factor: 3.748

  4 in total