| Literature DB >> 20553867 |
Charles J Choi1, Alysia R Belobraydich, Leo L Chan, Patrick C Mathias, Brian T Cunningham.
Abstract
Using both experimental assays and fluid-dynamic finite element simulation models, we directly compared the achievable performance limits of four distinct assay configurations for label-free detection of an analyte from a test sample on a biosensor surface. The assay configurations studied in this work included a biosensor incorporated into the bottom surface of a microplate well and a microfluidic channel. For each configuration, we compared assay performance for the scenario in which the entire bottom surface of the fluid-handling vessel is coated with capture ligands with assay performance for the scenario in which the capture ligands are applied in the form of localized spots. As a model system, we used detection of the protein biomarker tumor necrosis factor-alpha (TNF-alpha) using immobilized TNF-alpha capture antibody. Results show that the microfluidic assay format dramatically reduces the time required to establish a stable equilibrium. Spot-based assays are advantageous for microplate-based detection for reducing the time required for equilibrium sensor response. The results derived are generally applicable to any label-free biosensor technology and any ligand-analyte system with adjustable variables that include sensor mass density sensitivity, analyte-ligand adsorption/desorption rate constants, immobilized ligand density, flow channel geometry, flow rate, and spot size. Copyright 2010 Elsevier Inc. All rights reserved.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20553867 DOI: 10.1016/j.ab.2010.06.009
Source DB: PubMed Journal: Anal Biochem ISSN: 0003-2697 Impact factor: 3.365