Core temperature measurement in the intensive care unit.

D M Nierman1.   

Abstract

OBJECTIVE: To compare three devices that measure core body temperature at the bedside in ICU patients.
DESIGN: Prospective, consecutive sample.
SETTING: Voluntary community teaching hospital. PATIENTS: Fifteen patients, 78 +/- 6 (SD) yrs of age, admitted to the medical ICU over a 5-month period who had pulmonary artery catheters inserted as part of their routine care were studied. Thirteen patients were studied once, one patient twice, and one patient six times for a total of 21 sets of measurements.
INTERVENTIONS: All patients had urinary bladder thermistor catheters inserted just before pulmonary artery catheterization. Simultaneous core temperatures were measured for the duration of pulmonary artery catheterization every 4 hrs by the pulmonary artery thermistor catheter, the bladder thermistor catheter, and by a tympanic membrane infrared probe set on its core temperature setting. The three devices were then compared with each other in vitro using a specialized constant water bath setup. Finally, two of the tympanic membrane infrared probes were compared with each other in 20 ambulatory emergency department patients.
MEASUREMENTS AND MAIN RESULTS: Over 32 hrs of pulmonary artery catheterization, the pulmonary artery thermistor catheters and bladder thermistor catheters showed excellent agreement, with a bias of only -0.04 degrees C between the two. However, the bias comparing the tympanic membrane infrared probe with the pulmonary artery thermistor catheter was -0.38 degrees C, and the bias was -0.34 degrees C comparing the tympanic membrane infrared probe with the bladder thermistor catheter. The tympanic membrane infrared probe readings remained significantly higher than the pulmonary artery thermistor catheter or bladder thermistor catheter readings over the entire 32-hr period. The two tympanic membrane infrared probes compared with each other in emergency room patients showed excellent agreement (p less than .001). In the in vivo water bath setup, the tympanic membrane infrared probe on most of its settings again registered significantly (p less than .01) higher than both the pulmonary artery thermistor catheter and the bladder thermistor catheter.
CONCLUSIONS: Pulmonary artery thermistor catheters and bladder thermistor catheters appear to give consistent, highly reliable bedside measurements of core body temperature in ICU patients. The currently available device that measures core body temperature from the tympanic membrane appears to give erroneously high readings, and should be used with caution.

Entities:  

Mesh:

Year:  1991        PMID: 2055060     DOI: 10.1097/00003246-199106000-00015

Source DB:  PubMed          Journal:  Crit Care Med        ISSN: 0090-3493            Impact factor:   7.598


  12 in total

1.  Effect of a convenient single 90-mg pamidronate dose on biochemical markers of bone metabolism in patients with acute spinal cord injury.

Authors:  Jeffrey I Mechanick; Kan Liu; David M Nierman; Adam Stein
Journal:  J Spinal Cord Med       Date:  2006       Impact factor: 1.985

Review 2.  Thermometry in paediatric practice.

Authors:  A S El-Radhi; W Barry
Journal:  Arch Dis Child       Date:  2006-04       Impact factor: 3.791

3.  Comparison of temporal artery, rectal and esophageal core temperatures in children: Results of a pilot study.

Authors:  Fahad Al-Mukhaizeem; Upton Allen; Luba Komar; Basem Naser; Larry Roy; Derek Stephens; Stanley Read; Christina Kim; Suzanne Schuh
Journal:  Paediatr Child Health       Date:  2004-09       Impact factor: 2.253

4.  Effect of intravenous fluid warming on core body temperature during elective orthopedic procedures.

Authors:  Robert B E Brady; William T Poppell
Journal:  Can Vet J       Date:  2020-10       Impact factor: 1.008

5.  Temperature measurement in intensive care patients: comparison of urinary bladder, oesophageal, rectal, axillary, and inguinal methods versus pulmonary artery core method.

Authors:  J-Y Lefrant; L Muller; J Emmanuel de La Coussaye; M Benbabaali; C Lebris; N Zeitoun; C Mari; G Saïssi; J Ripart; J-J Eledjam
Journal:  Intensive Care Med       Date:  2003-02-08       Impact factor: 17.440

6.  Temporal Artery versus Bladder Thermometry during Adult Medical-Surgical Intensive Care Monitoring: An Observational Study.

Authors:  Henry T Stelfox; Sharon E Straus; William A Ghali; John Conly; Kevin Laupland; Adriane Lewin
Journal:  BMC Anesthesiol       Date:  2010-08-12       Impact factor: 2.217

7.  Altered circadian rhythmicity in patients in the ICU.

Authors:  Joost A C Gazendam; Hans P A Van Dongen; Devon A Grant; Neil S Freedman; Jan H Zwaveling; Richard J Schwab
Journal:  Chest       Date:  2013-08       Impact factor: 9.410

8.  Comparison of noncontact infrared thermometry and 3 commercial subcutaneous temperature transponding microchips with rectal thermometry in rhesus macaques (Macaca mulatta).

Authors:  Marla K Brunell
Journal:  J Am Assoc Lab Anim Sci       Date:  2012-07       Impact factor: 1.232

9.  Accuracy of tympanic temperature measurement using an infrared tympanic membrane thermometer.

Authors:  Gasim I Gasim; Imad R Musa; Mohamed T Abdien; Ishag Adam
Journal:  BMC Res Notes       Date:  2013-05-10

Review 10.  A recommended early goal-directed management guideline for the prevention of hypothermia-related transfusion, morbidity, and mortality in severely injured trauma patients.

Authors:  Ryan Perlman; Jeannie Callum; Claude Laflamme; Homer Tien; Barto Nascimento; Andrew Beckett; Asim Alam
Journal:  Crit Care       Date:  2016-04-20       Impact factor: 9.097