Influence of uncertainties in the diameter and refractive index of calibration polystyrene beads on the retrieval of aerosol optical properties using cavity ring down spectroscopy.

Rachael E H Miles1, Svemir Rudić, Andrew J Orr-Ewing, Jonathan P Reid.   

Abstract

We consider the impact of uncertainties in the refractive index and size of polystyrene beads on the retrieved optical properties of aerosol particles by aerosol cavity ring down spectroscopy (A-CRDS). Polystyrene beads are frequently used to verify and calibrate light extinction measurements by cavity ring down instruments. Any uncertainties in either the polymer particle size or the refractive index can contribute to systematic errors in properties retrieved for any subsequent measurements on other aerosol types. We demonstrate that the tolerances on bead size reported by the manufacturers can lead to a range in the real part of the refractive index of the polymer beads retrieved from A-CRDS measurements of as large as 2.9%. Further, we show that the current uncertainty in the refractive index of polystyrene beads in the visible part of the electromagnetic spectrum limits the accuracy with which the real part of the refractive index of other aerosol types can be retrieved to uncertainties of -0.5% and +0.3% at a minimum. This error should be included in any subsequent retrieval of aerosol optical properties from aerosol cavity ring down instruments that are dependent on polystyrene bead calibration. It is expected that such calibrations could lead to significantly larger uncertainties if the complex part of the refractive index is to be retrieved.

Entities:  

Year:  2010        PMID: 20545374     DOI: 10.1021/jp103246t

Source DB:  PubMed          Journal:  J Phys Chem A        ISSN: 1089-5639            Impact factor:   2.781


  3 in total

1.  Determination of refractive index, size, and solid content of monodisperse polystyrene microsphere suspensions for the characterization of optical phantoms.

Authors:  Peter Naglič; Yevhen Zelinskyi; Boštjan Likar; Miran Bürmen
Journal:  Biomed Opt Express       Date:  2020-03-11       Impact factor: 3.732

2.  Core and surface microgel mechanics are differentially sensitive to alternative crosslinking concentrations.

Authors:  Himansu Mohapatra; Terra M Kruger; Thiranjeewa I Lansakara; Alexei V Tivanski; Lewis L Stevens
Journal:  Soft Matter       Date:  2017-08-30       Impact factor: 3.679

Review 3.  Accurate Measurement of the Optical Properties of Single Aerosol Particles Using Cavity Ring-Down Spectroscopy.

Authors:  M I Cotterell; J W Knight; J P Reid; A J Orr-Ewing
Journal:  J Phys Chem A       Date:  2022-04-25       Impact factor: 2.944

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.