Literature DB >> 20542478

Development and validation of a sensitive ultra performance liquid chromatography tandem mass spectrometry method for the analysis of fentanyl and its major metabolite norfentanyl in urine and whole blood in forensic context.

Ruth Verplaetse1, Jan Tytgat.   

Abstract

Fentanyl and its major metabolite norfentanyl often occur in low doses in biological samples. Therefore, a highly sensitive liquid chromatography tandem mass spectrometry (LC-MS/MS) method has been developed and fully validated. Sample preparation was performed on a mixed-mode cation exchange solid phase extraction (SPE) cartridge with an additional alkaline wash step to decrease matrix effects and thus increase sensitivity. Ionization of fentanyl and norfentanyl with electrospray ionization (ESI) was more efficient than atmospheric pressure chemical ionization (APCI). The use of a mobile phase of high pH resulted in higher ESI signals than the conventional low pH mobile phases. In the final method, gradient elution with 10 mM ammonium bicarbonate (pH 9) and methanol was performed. A comparison of columns with different internal diameter and/or smaller particles showed optimal resolution and sensitivity when an Acquity C18 column (1.7 microm, 2.1 mm x 50 mm) was used. Deuterium labeled internal standards were used, but with careful evaluation of their stability since loss of deuteriums was seen. With limits of detection of 0.25 pg/ml for fentanyl and 2.5 pg/ml for norfentanyl in urine and 5 pg/ml for fentanyl and norfentanyl in whole blood the presented method is highly appropriate for the analysis of fentanyl and norfentanyl in forensic urine and blood samples. 2010 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20542478     DOI: 10.1016/j.jchromb.2010.05.033

Source DB:  PubMed          Journal:  J Chromatogr B Analyt Technol Biomed Life Sci        ISSN: 1570-0232            Impact factor:   3.205


  3 in total

1.  Determination of bisphenol-A levels in human amniotic fluid samples by liquid chromatography coupled with mass spectrometry.

Authors:  Mei Chen; Andrea G Edlow; Tao Lin; Nicole A Smith; Thomas F McElrath; Chensheng Lu
Journal:  J Sep Sci       Date:  2011-06-24       Impact factor: 3.645

2.  Quantitative low-volume assay for simultaneous determination of fentanyl, norfentanyl, and minor metabolites in human plasma and urine by liquid chromatography-tandem mass spectrometry (LC-MS/MS).

Authors:  Nina Sophia Mahlke; Victoria Ziesenitz; Gerd Mikus; Gisela Skopp
Journal:  Int J Legal Med       Date:  2014-07-06       Impact factor: 2.686

3.  LC-MS/MS-Based Method for the Multiplex Detection of 24 Fentanyl Analogues and Metabolites in Whole Blood at Sub ng mL-1 Concentrations.

Authors:  Kraig E Strayer; Heather M Antonides; Matthew P Juhascik; Raminta Daniulaityte; Ioana E Sizemore
Journal:  ACS Omega       Date:  2018-01-17