Numerous methods like distillation followed by iodometric titrations, gas chromatograph (GC)-flame ionization detector, gas chromatograph-mass spectrophotometer, GC-Headspace, Breath analyzer, and biosensors including alcohol dehydrogenase (enzymatic) have been used to determine blood alcohol concentration (BAC). In the present study, horizontal attenuated total reflectance-Fourier transform infrared spectroscopy had been used to determine BAC in whole blood. The asymmetric stretching frequency of C-C-O group of ethanol in water (1,045 cm(-1)) had been used to calculate BAC using Beer's Law. A seven-point calibration curve of ethanol was drawn in the concentration range 24-790 mg dL(-1). The curve showed good linearity over the concentration range used (r(2)=0.999, standard deviation=0.0023). The method is accurate, reproducible, rapid, simple, and nondestructive in nature. 2010 Elsevier Inc. All rights reserved.Entities: