Literature DB >> 20535751

Rapid monitoring of arrays of amino acids in clinical samples using capillary electrophoresis with contactless conductivity detection.

Petr Tůma1, Klára Málková, Eva Samcová, Karel Stulík.   

Abstract

CE with contactless conductivity detection has been used to separate 28 biogenic amino acids in a short capillary with an effective length of 18 cm. All the tested amino acids can be mutually separated in 0.5-10 mol/L acetic acid electrolytes. The time of analysis does not exceed 6 min and the LODs vary from 0.1 to 1.7 micromol/L. The CVs lie within the intervals 0.01-0.4% and 0.9-4% for the migration times and the analyte peak areas, respectively. The procedure has been successfully applied to the determinations of the whole amino acid spectra in blood plasma, urine, saliva and cerebrospinal fluid samples.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20535751     DOI: 10.1002/jssc.201000137

Source DB:  PubMed          Journal:  J Sep Sci        ISSN: 1615-9306            Impact factor:   3.645


  2 in total

1.  Mitochondrial Probe Methyltriphenylphosphonium (TPMP) Inhibits the Krebs Cycle Enzyme 2-Oxoglutarate Dehydrogenase.

Authors:  Moustafa Elkalaf; Petr Tůma; Martin Weiszenstein; Jan Polák; Jan Trnka
Journal:  PLoS One       Date:  2016-08-18       Impact factor: 3.240

2.  Quantitation of underivatized branched-chain amino acids in sport nutritional supplements by capillary electrophoresis with direct or indirect UV absorbance detection.

Authors:  Jun Qiu; Jinhao Wang; Zhongqi Xu; Huiqing Liu; Jie Ren
Journal:  PLoS One       Date:  2017-06-22       Impact factor: 3.240

  2 in total