Effects of variation of MRI parameters on signal homogeneity: a qualitative analysis for ferrous benzoic xylenon orange gel.

Aalia Nazir1, Muhammad Afzal, Saeed Ahmad Buzdar.   

Abstract

OBJECTIVE: This study focuses on the diverse effects of MRI parameters on image quality using widely available imaging pulse sequences.
METHODS: A tissue equivalent gel system has been used for Magnetic Resonance Imaging by using Xylenol orange dye with Fricke-Benzoic solution which was formulated by Kelly R.G (1998). The gel system was radiated while using 6MV photons from Varian Clinic Linear Accelerator.
RESULTS: The qualitative analysis include the effect of TR and TE on signal nonuniformity according to which the escalating repetition time gives a uniform signal having the average values calculated are 0.76%, 0.83%, 1.43% and 1.89% for the conventional spin echo (CSE), fast spin echo (FSE), gradient recalled echo (GRE) and fluid attenuated inversion recovery (FLAIR) respectively. FLAIR showed contradictory results due to longitudinal relaxation of the signal. This is because, at the longer value of inversion time (TI), the signal from simple fluids is nulled, thus reducing signal intensity.
CONCLUSION: The evaluation of signal uniformity for different pulse sequences demonstrates that the repetition time (TR) affects the signal homogeneity as it maintains image quality for CSE and FSE. However, a careful selection is required for FLAIR due to its sensitive behaviour for image uniformity.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20527646

Source DB:  PubMed          Journal:  J Pak Med Assoc        ISSN: 0030-9982            Impact factor:   0.781


  1 in total

1.  Does nitrogen gas bubbled through a low density polymer gel dosimeter solution affect the polymerization process?

Authors:  Daryoush Shahbazi-Gahrouei; Mehrdad Gholami; Tayyeb Allahverdi Pourfallah; Mohammad Keshtkar
Journal:  Adv Biomed Res       Date:  2015-05-11
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.