A pitfall of fractional flow reserve measurement.

Omar Ait Mokhtar1, Sebastien Armero, Laurent Bonello.   

Abstract

Fractional flow reserve is a simple and efficient tool to assess the severity of an intermediate lesion in order to determine the optimal therapy. However there are some limitations to its use. We observed that in patients with an occluded artery, FFR measurements in the vessel supplying collaterals can be underestimated leading to inappropriate therapy.

Entities:  

Mesh:

Year:  2010        PMID: 20516516

Source DB:  PubMed          Journal:  J Invasive Cardiol        ISSN: 1042-3931            Impact factor:   2.022


  1 in total

Review 1.  Clinical Application of Fractional Flow Reserve-Guided Percutaneous Coronary Intervention for Stable Coronary Artery Disease.

Authors:  Valay Parikh; Kanishk Agnihotri; Sabeeda Kadavath; Nileshkumar J Patel; J Dawn Abbott
Journal:  Curr Cardiol Rep       Date:  2016-04       Impact factor: 2.931