Fully hydroxyapatite-coated total hip replacement: ten-year results.

Màrius Valera Pertegàs1, Pedro Vergara-Valladolid, Xavier Crusi-Sererols, Rogelio Sancho-Navarro.   

Abstract

We assess 254 consecutive primary total hip replacements (THRs) using a fully hydroxyapatite (HA)-coated stem and socket (Furlong; JRI, London, UK). Minimum follow-up was 10 years. Harris hip score (HHS) increased from 49.8 preoperatively to 86.2 at follow-up. The most remarkable clinical finding was the near absence of thigh pain. After 10 years of follow-up, all stems met stability criteria and none needed revision for aseptic loosening. Three cups (1.2%) were revised for aseptic loosening. Two more cups were revised for recidivant dislocation. Varus placement of the femoral stem was present in 63 hips (24.8%) and valgus in 13 hips (5%). None of these technical defects showed radiologic progression, and there was no measurable correlation between them and the clinical score. Intraoperative femoral fractures occurred in 25 cases (9.8%) with no further influence on implant evolution. These findings strongly support the use of the JRI Furlong system for primary THR because it provides excellent clinical and radiologic results, with 0% and 1.2% cup and stem aseptic loosening, respectively, at 10 years. In addition, its tolerance to technical implantation defects makes it suitable even for less experienced hip surgeons.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20512777     DOI: 10.5301/HIP.2010.644

Source DB:  PubMed          Journal:  Hip Int        ISSN: 1120-7000            Impact factor:   2.135


  1 in total

1.  Influence of femoral broach shape on stem alignment using anterior approach for total hip arthroplasty: A radiologic comparative study of 3 different stems.

Authors:  Cécile Batailler; Camdon Fary; Elvire Servien; Sébastien Lustig
Journal:  PLoS One       Date:  2018-10-05       Impact factor: 3.240

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.