Literature DB >> 20512565

Alignment of retention time obtained from multicapillary column gas chromatography used for VOC analysis with ion mobility spectrometry.

Thorsten Perl1, Bertram Bödeker, Melanie Jünger, Jürgen Nolte, Wolfgang Vautz.   

Abstract

Multicapillary column (MCC) ion mobility spectrometers (IMS) are increasingly in demand for medical diagnosis, biological applications and process control. In a MCC-IMS, volatile compounds are differentiated by specific retention time and ion mobility when rapid preseparation techniques are applied, e.g. for the analysis of complex and humid samples. Therefore, high accuracy in the determination of both parameters is required for reliable identification of the signals. The retention time in the MCC is the subject of the present investigation because, for such columns, small deviations in temperature and flow velocity may cause significant changes in retention time. Therefore, a universal correction procedure would be a helpful tool to increase the accuracy of the data obtained from a gas-chromatographic preseparation. Although the effect of the carrier gas flow velocity and temperature on retention time is not linear, it could be demonstrated that a linear alignment can compensate for the changes in retention time due to common minor deviations of both the carrier gas flow velocity and the column temperature around the MCC-IMS standard operation conditions. Therefore, an effective linear alignment procedure for the correction of those deviations has been developed from the analyses of defined gas mixtures under various experimental conditions. This procedure was then applied to data sets generated from real breath analyses obtained in clinical studies using different instruments at different measuring sites for validation. The variation in the retention time of known signals, especially for compounds with higher retention times, was significantly improved. The alignment of the retention time--an indispensable procedure to achieve a more precise identification of analytes--using the proposed method reduces the random error caused by small accidental deviations in column temperature and flow velocity significantly.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20512565      PMCID: PMC2895891          DOI: 10.1007/s00216-010-3798-1

Source DB:  PubMed          Journal:  Anal Bioanal Chem        ISSN: 1618-2642            Impact factor:   4.142


  9 in total

Review 1.  Ion mobility spectrometry: arriving on site and moving beyond a low profile.

Authors:  J I Baumbach; G A Eiceman
Journal:  Appl Spectrosc       Date:  1999-09       Impact factor: 2.388

2.  Detection of human metabolites using multi-capillary columns coupled to ion mobility spectrometers.

Authors:  Vera Ruzsanyi; Jörg Ingo Baumbach; Stefanie Sielemann; P Litterst; M Westhoff; Lutz Freitag
Journal:  J Chromatogr A       Date:  2005-08-19       Impact factor: 4.759

3.  Ion mobility spectrometry for food quality and safety.

Authors:  W Vautz; D Zimmermann; M Hartmann; J I Baumbach; J Nolte; J Jung
Journal:  Food Addit Contam       Date:  2006-11

Review 4.  Ion mobility spectrometry detection for gas chromatography.

Authors:  Abu B Kanu; Herbert H Hill
Journal:  J Chromatogr A       Date:  2007-11-12       Impact factor: 4.759

5.  Ion mobility spectrometry in the diagnosis of sarcoidosis: results of a feasibility study.

Authors:  M Westhoff; P Litterst; L Freitag; J I Baumbach
Journal:  J Physiol Pharmacol       Date:  2007-11       Impact factor: 3.011

Review 6.  Ion mobility spectrometry.

Authors:  H H Hill; W F Siems; R H St Louis; D G McMinn
Journal:  Anal Chem       Date:  1990-12-01       Impact factor: 6.986

7.  Determination of serum propofol concentrations by breath analysis using ion mobility spectrometry.

Authors:  T Perl; E Carstens; A Hirn; M Quintel; W Vautz; J Nolte; M Jünger
Journal:  Br J Anaesth       Date:  2009-11-03       Impact factor: 9.166

8.  Microfabricated differential mobility spectrometry with pyrolysis gas chromatography for chemical characterization of bacteria.

Authors:  Hartwig Schmidt; Fathulla Tadjimukhamedov; Isabelle V Mohrenz; Geoffrey B Smith; Gary A Eiceman
Journal:  Anal Chem       Date:  2004-09-01       Impact factor: 6.986

9.  Ion mobility spectrometry for the detection of volatile organic compounds in exhaled breath of patients with lung cancer: results of a pilot study.

Authors:  M Westhoff; P Litterst; L Freitag; W Urfer; S Bader; J-I Baumbach
Journal:  Thorax       Date:  2009-01-21       Impact factor: 9.139

  9 in total
  6 in total

Review 1.  Review on ion mobility spectrometry. Part 2: hyphenated methods and effects of experimental parameters.

Authors:  R Cumeras; E Figueras; C E Davis; J I Baumbach; I Gràcia
Journal:  Analyst       Date:  2015-03-07       Impact factor: 4.616

2.  Ion mobility spectrometry for microbial volatile organic compounds: a new identification tool for human pathogenic bacteria.

Authors:  Melanie Jünger; Wolfgang Vautz; Martin Kuhns; Lena Hofmann; Siobhán Ulbricht; Jörg Ingo Baumbach; Michael Quintel; Thorsten Perl
Journal:  Appl Microbiol Biotechnol       Date:  2012-02-12       Impact factor: 4.813

3.  Computational methods for metabolomic data analysis of ion mobility spectrometry data-reviewing the state of the art.

Authors:  Anne-Christin Hauschild; Till Schneider; Josch Pauling; Kathrin Rupp; Mi Jang; Jörg Ingo Baumbach; Jan Baumbach
Journal:  Metabolites       Date:  2012-10-16

Review 4.  Assessment, origin, and implementation of breath volatile cancer markers.

Authors:  Hossam Haick; Yoav Y Broza; Pawel Mochalski; Vera Ruzsanyi; Anton Amann
Journal:  Chem Soc Rev       Date:  2013-12-04       Impact factor: 54.564

5.  Detection and validation of volatile metabolic patterns over different strains of two human pathogenic bacteria during their growth in a complex medium using multi-capillary column-ion mobility spectrometry (MCC-IMS).

Authors:  Nils Kunze; Julia Göpel; Martin Kuhns; Melanie Jünger; Michael Quintel; Thorsten Perl
Journal:  Appl Microbiol Biotechnol       Date:  2013-03-07       Impact factor: 4.813

6.  Smell the change: On the potential of gas-chromatographic ion mobility spectrometry in ecosystem monitoring.

Authors:  Wolfgang Vautz; Chandrasekhara Hariharan; Maximilian Weigend
Journal:  Ecol Evol       Date:  2018-04-02       Impact factor: 2.912

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.