Literature DB >> 20506940

Acetabular screw head-induced ceramic acetabular liner fracture in cementless ceramic-on-ceramic total hip arthroplasty.

Su Chan Lee1, Kwang Am Jung, Chang Hyun Nam, Tea Ho Kim, Nong Kyoum Ahn, Seung Hyun Hwang.   

Abstract

Ceramic liner fractures are rare after ceramic-on-ceramic THA. This article describes a case of an early ceramic liner fracture caused by impingement with a tilted acetabular screw head 2 months after cementless ceramic-on-ceramic THA. A 59-year-old man underwent primary THA for avascular necrosis of his right femoral head. The implant used was an Osteonics Secur-Fit HA ceramic-on-ceramic bearing system. The metal shell was fixed with 1 cancellous bone screw. During the index procedure, the acetabular screw seemed tilted in the metal shell hole, but the liner was fully seated in the metal shell without difficulty. However, 2 months later he re-presented due to pain and crepitation in the right hip. Radiographs showed that the ceramic acetabular insert had fractured. At revision, ceramic insert had fractured into large and numerous comminuted fragments, and the acetabular screw head was slightly tilted and protruded over the inner surface of the metal shell, which had worn eccentrically. The ceramic inner head and metal shell were visibly intact. Because the metal shell-bone fixation was firm, a new identical design ceramic liner and head were fitted, and no adverse event has occurred since. This case suggests that a complete check of the inside of the shell should be made when using an acetabular screw and ceramic liner. In particular, the screw head must not be tilted or left proud of the inside surface of the shell. Correct acetabular screw direction and seating are essential to avoid detrimental clinical consequences. Copyright 2010, SLACK Incorporated.

Entities:  

Mesh:

Year:  2010        PMID: 20506940     DOI: 10.3928/01477447-20100329-30

Source DB:  PubMed          Journal:  Orthopedics        ISSN: 0147-7447            Impact factor:   1.390


  3 in total

Review 1.  Strategies for head and inlay exchange in revision hip arthroplasty.

Authors:  Martin Pospischill; Karl Knahr
Journal:  Int Orthop       Date:  2010-11-19       Impact factor: 3.075

2.  Extraction of a well fixed but fractured ceramic acetabular liner.

Authors:  David Ferguson; Robert Metcalf
Journal:  Arthroplast Today       Date:  2015-03-20

Review 3.  Fracture of ceramic bearing surfaces following total hip replacement: a systematic review.

Authors:  Francesco Traina; Marcello De Fine; Alberto Di Martino; Cesare Faldini
Journal:  Biomed Res Int       Date:  2013-06-13       Impact factor: 3.411