External nasal valve collapse - a case-control and interventional study employing a novel internal nasal dilator (Nasanita).

H Riechelmann1, E Karow, D DiDio, F Kral.   

Abstract

BACKGROUND: Nasal alar collapse is a common problem and difficult to assess and treat.
METHODS: In 10 healthy controls and 10 patients with alar collapse, the size of the external nasal valve was analyzed on standardized nasal base photographs during quiet breathing and forced inspiration. A novel internal nasal dilator (Nasanita, Siemens & Co, Germany) was employed to assess the effects of a therapeutic intervention. In addition, active anterior rhinomanometry was performed.
RESULTS: During quiet breathing, the external nasal valves were significantly smaller in patients with alar collapse (0.3 +/- 0.08 cm2) than in controls (0.7 +/- 0.2 cm2; p < 0.001). In heal-thy controls, forced inspiration did not significantly alter the size of the external nasal valve (-1.8% +/- 27.5%; p = 0.84), whereas it significantly decreased the external valve area in patients with alar collapse (-42.1% +/- 26.4%; p = 0.001). The internal nasal dilator significantly increased external valve areas during quiet breathing and forced inspiration and completely abolished alar collapse. Nasal airflow at a transnasal pressure difference of 150 Pa was not correlated with external valve size. Nasal airflow increased significantly after inserting the internal nasal dilator to 1300 +/- 370 ml/s (p < 0.001) in controls and 1300 +/-300 ml/s (p < 0.01) in patients.
CONCLUSION: A small sized external nasal valve appears to be a major causative factor of alar collapse. A novel internal nasal dilator effectively enlarged the external nasal valve, abolished alar collapse and improved nasal airflow.

Entities:  

Mesh:

Year:  2010        PMID: 20502758     DOI: 10.4193/Rhin09.121

Source DB:  PubMed          Journal:  Rhinology        ISSN: 0300-0729            Impact factor:   3.681


  2 in total

1.  [Treatment of nasal valve stenosis].

Authors:  W Heppt; T Hildebrandt; J Vent
Journal:  HNO       Date:  2015-03       Impact factor: 1.284

2.  Displaced nasal dilator caused severe pain: Case report and literature review.

Authors:  Eva M Jungmark; Eva K Ellegård
Journal:  Allergy Rhinol (Providence)       Date:  2012-09-18
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.