R Kapil1, S Dhawan, Bhupinder Singh.
Abstract
A rapid, sensitive, simple, and cost-effective spectrofluorimetric method was developed for the estimation of rivastigmine in bulk and pharmaceutical formulations. The relative fluorescence intensity of rivastigmine was measured in triple distilled water at an excitation wavelength of 220 nm and an emission wavelength of 289 nm. Linearity range was found to be 100 to 4000 ng/ml. The method was validated for various parameters as per the ICH guidelines and USP requirements. The detection and quantitation limits were found to be 20.5 and 62.1 ng/ml, respectively. The results demonstrate that the procedure is accurate, precise, and reproducible, while being simple and rapid too. The results were found to be in good agreement with the label claims.Entities:
Keywords: Detection limit; fluorescence spectrophotometry; fluorimetry; linearity; quantitation limit; validation
Year: 2009 PMID: 20502586 PMCID: PMC2866359 DOI: 10.4103/0250-474X.58179
Source DB: PubMed Journal: Indian J Pharm Sci ISSN: 0250-474X Impact factor: 0.975
Fig. 1Emission scan of rivastigmine at the excitation wavelength of 220 nm
CALIBRATION DATA OF RIVASTIGMINE
| Drug concentration(ng/ml) | Mean relative fluorescence intensity (±SD) | % RSD | Predicted drug concentration (ng/ml) |
|---|---|---|---|
| 100 | 47.3±0.66 | 1.405 | 100.8 |
| 150 | 65.7±0.62 | 0.946 | 146.7 |
| 200 | 85.3±0.69 | 0.812 | 195.4 |
| 250 | 105.9±3.02 | 2.891 | 246.7 |
| 300 | 122.1±4.89 | 3.995 | 287.0 |
| 400 | 172.8±5.09 | 2.946 | 413.2 |
| 500 | 202.6±4.52 | 2.233 | 487.4 |
| 1000 | 400.1±7.00 | 1.749 | 979.1 |
| 2000 | 812.8±5.79 | 0.713 | 2006.4 |
| 2500 | 1021.1±7.78 | 0.761 | 2524.7 |
| 3000 | 1217.9±7.07 | 0.580 | 3015.2 |
| 4000 | 1604.2±7.07 | 0.440 | 3976.1 |
Fig. 2Validation of linearity of the analytical method of rivastigmine
ACCURACY DATA FOR THE DEVELOPED ANALYTICAL METHOD
| Level | Predicted concentration (ng/ml) | Mean % recovery (± SD) | Bias (%) | ||
|---|---|---|---|---|---|
| Range | Mean (± SD) | % RSD | |||
| LQC | 148.0-150.9 | 149.7±1.47 | 0.983 | 99.82±0.98 | − 0.18 |
| MQC | 494.2-503.1 | 500.3±4.40 | 0.880 | 100.06±0.88 | 0.06 |
| HQC | 2495.3-2512.2 | 2504.3±8.29 | 0.331 | 100.16±0.33 | 0.16 |
DATA DEPICTING INTERMEDIATE PRECISION STUDY
| Level | Intra-day repeatability, % RSD (n = 3) | Inter-day repeatability, % RSD (n = 18) | |||
|---|---|---|---|---|---|
| Analyst | Day 1 | Day 2 | Day 3 | ||
| LQC | 1 | 0.833 | 0.784 | 0.973 | 0.971 |
| 2 | 0.925 | 0.902 | 0.663 | ||
| MQC | 1 | 0.529 | 0.564 | 0.871 | 0.762 |
| 2 | 0.806 | 0.781 | 0.691 | ||
| HQC | 1 | 0.410 | 0.448 | 0.487 | 0.396 |
| 2 | 0.323 | 0.215 | 0.428 | ||