Effect of inhomogeneities and source position on dose distribution of nucletron high dose rate Ir-192 brachytherapy source by Monte Carlo simulation.

R M Chandola1, S Tiwari, M K Kowar, V Choudhary.   

Abstract

BACKGROUND: The presence of least dense dry air and highly dense cortical bone in the path of radiation and the position of source, near or far from the surface of patient, affects the exact dose delivery like in breast brachytherapy. AIM: This study aims to find out the dose difference in the presence of inhomogeneities like cortical bone and dry air as well as to find out difference of dose due to position of source in water phantom of high dose rate (HDR) 192 Ir nucletron microselectron v2 (mHDRv2) brachytherapy source using Monte Carlo (MC) simulation EGSnrc code, so that the results could be used in Treatment Planning System (TPS) for more precise brachytherapy treatment. SETTINGS AND
DESIGN: The settings and design are done using different software of the computer. METHODS AND MATERIALS: For this study, the said source, water phantom of volume 30 x 30 x 30 cm 3, inhomogeneities each of volume 1 x 2 x 2 cm3 with their position, water of water phantom and position of source are modeled using three-dimensional MC EGSnrc code. STATISTICAL ANALYSIS USED: Mean and probability are used for results and discussion.
RESULTS: The % relative dose difference is calculated here as 5.5 to 6.5% higher and 4.5 to 5% lower in the presence of air and cortical bone respectively at transverse axis of the source, which may be due to difference of linear attenuation coefficients of the inhomogeneities. However, when the source was positioned at 1 cm distance from the surface of water phantom, the near points between 1 to 2 cm and 3 to 8 cm. from the source, at its transverse axis, were 2 to 3.5% and 4 to 16% underdose to the dose when the source was positioned at mid-point of water phantom. This may be due to lack of back scatter material when the source was positioned very near to the surface of said water phantom and overlap of the additional cause of missing scatter component with the primary dose for near points from the source. These results were found in good agreement with literature data.
CONCLUSION: The results can be used in TPS.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20479548     DOI: 10.4103/0973-1482.63567

Source DB:  PubMed          Journal:  J Cancer Res Ther        ISSN: 1998-4138            Impact factor:   1.805


  5 in total

Review 1.  A review of dosimetric impact of implementation of model-based dose calculation algorithms (MBDCAs) for HDR brachytherapy.

Authors:  Yousif A M Yousif; Alexander F I Osman; Mohammed A Halato
Journal:  Phys Eng Sci Med       Date:  2021-06-17

2.  Effect of tissue composition on dose distribution in electron beam radiotherapy.

Authors:  M Ghorbani; Z S Tabatabaei; A Vejdani Noghreiyan; H Vosoughi; C Knaup
Journal:  J Biomed Phys Eng       Date:  2015-03-04

3.  Dose error from deviation of dwell time and source position for high dose-rate 192Ir in remote afterloading system.

Authors:  Hiroyuki Okamoto; Ako Aikawa; Akihisa Wakita; Kotaro Yoshio; Naoya Murakami; Satoshi Nakamura; Minoru Hamada; Yoshihisa Abe; Jun Itami
Journal:  J Radiat Res       Date:  2014-02-23       Impact factor: 2.724

4.  Effect of tissue composition on dose distribution in brachytherapy with various photon emitting sources.

Authors:  Mahdi Ghorbani; Fateme Salahshour; Abbas Haghparast; Toktam Ahmadi Moghaddas; Courtney Knaup
Journal:  J Contemp Brachytherapy       Date:  2014-04-03

5.  Dosimetric evaluation of multilumen intracavitary balloon applicator rotation in high-dose-rate brachytherapy for breast cancer.

Authors:  Yongbok Kim; Mark G Trombetta
Journal:  J Appl Clin Med Phys       Date:  2014-01-06       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.