High-sensitive fluorescence endoscope using electrocardiograph-synchronized multiple exposure.

Takehiro Ando1, Kazuhiro Taniguchi, Hongho Kim, Sanghyun Joung, Etsuko Kobayashi, Hongen Liao, Shunei Kyo, Ichiro Sakuma.   

Abstract

PURPOSE: Fluorescence-based measurement of cardiac disease, using autofluorescent substances that already exist in the heart, has not been used for endoscopic surgery because the endoscopic lenses cannot transmit sufficient light. A highly sensitive fluorescence endoscope using an electrocardiograph (ECG)-synchronized multiple exposure (ESME) approach was developed that provides a bright fluorescent image.
METHODS: A system was developed consisting of an endoscope, an excitation light, an ECG amplifier, a trigger and delay unit, and a computer. This system is based on periodic motion of the heart. Since the shape of the heart can be photographed by ECG triggering in a similar manner, a bright image can be synthesized by accumulating multiple trigger-captured images. Laboratory and in vivo experiments were performed to confirm the effectiveness of ESME.
RESULTS: The experimental results revealed that the trigger unit generated the synchronization signals required to produce high-quality images of the heart depending on heart rate. The difference among trigger-captured images from the actual organ, which affects the quality of ESME images, was estimated at 0.65 mm from the calculated displacement of a marker on the heart. The results also revealed that a bright fluorescent image can be captured by ESME.
CONCLUSION: A highly sensitive fluorescence endoscope using ESME was developed and successfully tested. The experimental results indicated that the method enabled high-quality image acquisition in a very low illumination environment. This system is effective for the observation of faint fluorescence in the heart and is useful for the intraoperative examination of the heart status.

Entities:  

Mesh:

Year:  2010        PMID: 20473575     DOI: 10.1007/s11548-010-0478-8

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  17 in total

1.  Noninvasive coronary angiography by retrospectively ECG-gated multislice spiral CT.

Authors:  S Achenbach; S Ulzheimer; U Baum; M Kachelriess; D Ropers; T Giesler; W Bautz; W G Daniel; W A Kalender; W Moshage
Journal:  Circulation       Date:  2000-12-05       Impact factor: 29.690

2.  Surgical navigation by autostereoscopic image overlay of integral videography.

Authors:  Hongen Liao; Nobuhiko Hata; Susumu Nakajima; Makoto Iwahara; Ichiro Sakuma; Takeyoshi Dohi
Journal:  IEEE Trans Inf Technol Biomed       Date:  2004-06

3.  Toward robotized beating heart TECABG: assessment of the heart dynamics using high-speed vision.

Authors:  Loïc Cuvillon; Jacques Gangloff; Michel de Mathelin; Antonello Forgione
Journal:  Med Image Comput Comput Assist Interv       Date:  2005

4.  Computer-enhanced "robotic" cardiac surgery: experience in 148 patients.

Authors:  F W Mohr; V Falk; A Diegeler; T Walther; J F Gummert; J Bucerius; S Jacobs; R Autschbach
Journal:  J Thorac Cardiovasc Surg       Date:  2001-05       Impact factor: 5.209

5.  Photodynamic therapy of human glioma spheroids using 5-aminolevulinic acid.

Authors:  S J Madsen; C H Sun; B J Tromberg; V P Wallace; H Hirschberg
Journal:  Photochem Photobiol       Date:  2000-07       Impact factor: 3.421

6.  Minimally invasive cardiac valve surgery improves patient satisfaction while reducing costs of cardiac valve replacement and repair.

Authors:  L H Cohn; D H Adams; G S Couper; D P Bichell; D M Rosborough; S P Sears; S F Aranki
Journal:  Ann Surg       Date:  1997-10       Impact factor: 12.969

7.  Improving the quality of coronary bypass surgery with intraoperative angiography: validation of a new technique.

Authors:  Nimesh D Desai; Senri Miwa; David Kodama; Gideon Cohen; George T Christakis; Bernard S Goldman; Mark O Baerlocher; Marc P Pelletier; Stephen E Fremes
Journal:  J Am Coll Cardiol       Date:  2005-09-28       Impact factor: 24.094

8.  Intraoperative optical spectroscopy identifies infiltrating glioma margins with high sensitivity.

Authors:  Steven A Toms; Wei-Chiang Lin; Robert J Weil; Mahlon D Johnson; E Duco Jansen; Anita Mahadevan-Jansen
Journal:  Neurosurgery       Date:  2005-10       Impact factor: 4.654

9.  Preliminary experience with a novel intraoperative fluorescence imaging technique to evaluate the patency of bypass grafts in total arterial revascularization.

Authors:  David P Taggart; Bikram Choudhary; Kyriakos Anastasiadis; Yasir Abu-Omar; Lognathen Balacumaraswami; David W Pigott
Journal:  Ann Thorac Surg       Date:  2003-03       Impact factor: 4.330

10.  The detection of atrial and ventricular septal defects with electrocardiographically synchronized magnetic resonance imaging.

Authors:  D G Lowell; D A Turner; S M Smith; G H Bucheleres; B A Santucci; R J Gresick; D O Monson
Journal:  Circulation       Date:  1986-01       Impact factor: 29.690

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.