R E Wood1, A M Harris, E J van der Merwe, C J Nortjé.
Abstract
A tissue-equivalent anthropomorphic human phantom was used with a lithium fluoride thermoluminescent dosimetry system to evaluate the radiation absorbed dose to the ovarian and testicular region during dental radiologic procedures. Measurements were made with and without personal lead shielding devices consisting of thyroid collar and apron of 0.25 mm lead thickness equivalence. The radiation absorbed dose with or without lead shielding did not differ significantly from control dosimeters in vertex occlusal and periapical views (p greater than 0.05). Personal lead shielding devices did reduce gonadal dose in the case of accidental exposure (p less than 0.05). A leaded apron of 0.25 mm lead thickness equivalent was permeable to radiation in direct exposure testing.Entities:
Mesh:
Substances:
Year: 1991 PMID: 2047109 DOI: 10.1016/0030-4220(91)90377-o
Source DB: PubMed Journal: Oral Surg Oral Med Oral Pathol ISSN: 0030-4220