HyperForm balloon remodeling in the endovascular treatment of anterior cerebral, middle cerebral, and anterior communicating artery aneurysms: clinical and angiographic follow-up results in 800 consecutive patients.

H Saruhan Cekirge1, Kivilcim Yavuz, Serdar Geyik, Isil Saatci.   

Abstract

OBJECT: The purpose of this paper was to present the safety, efficacy, and clinical/angiographic follow-up results of HyperForm balloon-assisted endosaccular coil occlusion of distal anterior circulation bifurcation aneurysms.
METHODS: Over a 7-year period, the authors treated 864 middle cerebral artery, distal anterior cerebral artery bifurcation, and anterior communicating artery aneurysms by means of coil embolization with HyperForm balloon assistance in 800 patients. In 37 aneurysms, 2 HyperForm balloons were used simultaneously for remodeling.
RESULTS: The overall mortality rate was 7.1%, including 1.4% procedural mortality. Various neurological deficits were present at discharge in 8.9% of the patients, and 4.4% had permanent disabling morbidity 6 months posttreatment (modified Rankin Scale score ≥ 2). Thromboembolic complications developed during the treatment of 15 aneurysms (1.7%) causing morbidity or mortality in 10 cases (1.3%). There were 14 intraoperative perforations (1.6%). In all 14 cases, the HyperForm balloon saved patients from severe bleeding. The perforation led to morbidity or mortality in 3 cases (0.4%); there were no negative consequences in 11. There were 726 patients with 757 aneurysms (87.6%) available for follow-up. Control angiograms were obtained at 6 months in 386 patients, at 1 year in 267, and at 2 years in 104, revealing an 82% complete obliteration rate according to the most recent follow-up angiograms.
CONCLUSIONS: The satisfactory results obtained in this experience demonstrate that HyperForm balloon remodeling provides strong benefits for the endovascular management of middle cerebral, anterior cerebral, and anterior communicating artery aneurysms without increasing the risk of treatment. Not only does this technique allow for the safe treatment of these aneurysms, but it also expands the indications of endovascular treatment to include aneurysms that otherwise cannot be treated with simple coil embolization.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20469985     DOI: 10.3171/2010.3.JNS081131

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  17 in total

1.  Early Experience with the TransForm™ Occlusion Balloon Catheter: A Single-Center Study.

Authors:  Syed A Quadri; Vivek Ramakrishnan; Omid Hariri; M Asif Taqi
Journal:  Interv Neurol       Date:  2015-06-24

2.  The place for remodeling technique and stenting in the endovascular management of intracranial aneurysms: a single-center analysis from 2008 to 2010.

Authors:  Laurent Pierot; Girish Rajpal; Krzysztof Kadziolka; Coralie Barbe
Journal:  Neuroradiology       Date:  2011-11-09       Impact factor: 2.804

Review 3.  "Coil mainly" policy in management of intracranial ACoA aneurysms: single-centre experience with the systematic review of literature and meta-analysis.

Authors:  Anna Steklacova; Ondrej Bradac; Patricia de Lacy; Jiri Lacman; Frantisek Charvat; Vladimir Benes
Journal:  Neurosurg Rev       Date:  2017-11-28       Impact factor: 3.042

4.  Balloon-assisted rapid intermittent sequential coiling (BRISC) technique for the treatment of complex wide-necked intracranial aneurysms.

Authors:  J Modi; M Eesa; B K Menon; J H Wong; M Goyal
Journal:  Interv Neuroradiol       Date:  2011-04-29       Impact factor: 1.610

5.  Endovascular Treatment of Wide-Necked Intracranial Aneurysms Using Balloon-Assisted Technique with HyperForm Balloon.

Authors:  Sang O Youn; Jae Il Lee; Jun Kyung Ko; Tae Hong Lee; Chang Hwa Choi
Journal:  J Korean Neurosurg Soc       Date:  2010-09-30

6.  Double stent-assisted coil embolization treatment for bifurcation aneurysms: immediate treatment results and long-term angiographic outcome.

Authors:  K Yavuz; S Geyik; S Cekirge; I Saatci
Journal:  AJNR Am J Neuroradiol       Date:  2013-03-28       Impact factor: 3.825

7.  Initial experience of intracranial aneurysm embolization using the balloon remodeling technique with Scepter C, a new double-lumen balloon.

Authors:  B Gory; I Kessler; G Seizem Nakiri; R Riva; M Al-Khawaldeh; C Mounayer
Journal:  Interv Neuroradiol       Date:  2012-09-10       Impact factor: 1.610

8.  TransForm occlusion balloon catheter for the treatment of intracranial aneurysms, initial experience.

Authors:  Bruno Bartolini; Raphaël Blanc; Silvia Pistocchi; Hocine Redjem; Gabriele Ciccio; Michel Piotin
Journal:  Interv Neuroradiol       Date:  2015-05-07       Impact factor: 1.610

9.  A Prospective Multicenter Trial of the TransForm Occlusion Balloon Catheter: Trial Design and Results.

Authors:  M Asif Taqi; Syed A Quadri; Ajit S Puri; Brian F Fitzsimmons; Jianhua Nancy Jin; Ansaar T Rai; Curtis A Given; Javier Masso; Ciaran J Powers; Joey D English; Osama O Zaidat
Journal:  Interv Neurol       Date:  2017-11-15

10.  Simple coiling using single or multiple catheters without balloons or stents in middle cerebral artery bifurcation aneurysms.

Authors:  Sung-Chul Jin; O-Ki Kwon; Chang Wan Oh; Jae Seung Bang; Gyojun Hwang; Nam Mi Park; Eun A Jung; Moon Hee Han; Hyun-Seung Kang; Hyun Park
Journal:  Neuroradiology       Date:  2012-11-29       Impact factor: 2.804

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.