| Literature DB >> 2046335 |
C N Ong1, G L Sia, S E Chia, W H Phoon, K T Tan.
Abstract
A simple and sensitive method for determining urinary cyclohexanol, the main metabolite of cyclohexanone, by hydrolysis and gas chromatography (GC) with a flame ionization detector was developed. A 2-mL urine sample was hydrolyzed with 0.4 mL of concentrated HCl and followed by extracting twice with diethylether. Two microL of the filtrate was injected into the GC with a methyl silicone column. The detection limit is estimated to be 0.4 mg/L. The coefficient of variation for the procedure is 8% and 10% for the range of concentration 5 and 50 mg/L, respectively. The within-run variation was 5.4% and between-day variation was 9.67%. The method was verified with urine samples collected from workers exposed to cyclohexanone. An excellent correlation (r = 0.88) was observed between environmental cyclohexanone exposure and cyclohexanol in urine. The procedure is relatively simple and reproducible and it can be applied for occupational health measurement of cyclohexanone exposure.Entities:
Mesh:
Substances:
Year: 1991 PMID: 2046335 DOI: 10.1093/jat/15.1.13