Literature DB >> 20459354

Evaluation of the TEST 1 erythrocyte sedimentation rate system and intra- and inter-laboratory quality control using new latex control materials.

Choong-Hwan Cha1, Young Joo Cha, Chan-Jeoung Park, Hyun Kyung Kim, Eun-Jong Cha, Duck Hee Kim, Jae-Seol Jung, Mi-Jung Kim, Seongsoo Jang, Hyun-Sook Chi, Dong Soon Lee.   

Abstract

BACKGROUND: The erythrocyte sedimentation rate (ESR) test has been considered to be a simple procedure, not requiring quality control (QC). However, QC is essential for accuracy and precision. We evaluated the TEST 1 ESR system and performed QC procedures using newly developed latex control materials in three hospitals.
METHODS: Using tripotassium ethylenediaminetetraacetic acid blood samples (n=184), we compared TEST 1 ESR values with Westergren ESR data and evaluated intra-assay precision. Three levels of latex control materials were used to assess inter-assay precision. Reference range assessment was done using samples from 220 healthy individuals. Inter-laboratory QC with latex control materials in three hospitals was performed.
RESULTS: Correlation between TEST 1 ESR and Westergren ESR results was good (p<0.001). Intra-assay precision [coefficients of variation (CV) 6.6%-21.7%] with patient samples and inter-assay precision (CV 0.0%-6.8%) with latex control materials were satisfactory. The reference ranges of 2-10 mm/h for males and 2-19 mm/h for females were established. Inter-laboratory QC data with latex control materials in three hospitals demonstrated good accuracy and satisfactory precision (CV 0.0%-14.4%).
CONCLUSIONS: Our results demonstrate that the TEST 1 QC is reliable and the latex control materials are valuable for inter-laboratory proficiency testing.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20459354     DOI: 10.1515/CCLM.2010.162

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  2 in total

1.  Test-1 analyzer and conventional Westergren method for erythrocyte sedimentation rate: A comparative study between two laboratories.

Authors:  Cigdem Sonmez; Ozlem Ceylan Dogan; Aysegul Ozturk Kaymak; Nedim Akkaya; Kadir Okhan Akin; Gulcan Guntas
Journal:  J Clin Lab Anal       Date:  2018-01-17       Impact factor: 2.352

2.  Erythrocyte sedimentation rate and hemoglobin-binding protein in free-living box turtles (Terrapene spp.).

Authors:  Laura Adamovicz; Sarah J Baker; Ethan Kessler; Marta Kelly; Samantha Johnson; John Winter; Christopher A Phillips; Matthew C Allender
Journal:  PLoS One       Date:  2020-06-17       Impact factor: 3.240

  2 in total