Comparison of the automated Phoenix with the Vitek 2 for the identification of Streptococcus pneumoniae.

Scott A Mittman1, Richard C Huard, Phyllis Della-Latta, Susan Whittier.   

Abstract

Rapid and accurate identification of Streptococcus pneumoniae is a critical component in the optimal management of infected patients. The performance of the BD Phoenix Automated Microbiology System (BD Diagnostic Systems, Sparks, Md.) was evaluated for identification of S. pneumoniae (n = 311) and was compared to the Vitek 2 (bioMérieux, Marcy l'Etoile, France). Strains with discordant identification between methods were resolved with 16S rRNA gene sequencing as the gold standard. The Phoenix and the Vitek 2 correctly identified 96.8% (n = 301) and 95.2% (n = 296) of S. pneumoniae strains, respectively. Overall, there was no statistically significant difference in the performance of the 2 automated systems for the identification of S. pneumoniae in this study. The Vitek 2 mean time-to-results for all streptococcal identification was 1.5 h faster than that for the Phoenix. We conclude that the automated Phoenix and the Vitek 2 systems are comparable in their ability to identify S. pneumoniae and are preferable to the use of routine biochemical assays, which have delayed time-to-results and are not dependably accurate.

Entities:  

Mesh:

Year:  2010        PMID: 20453899     DOI: 10.1139/w10-016

Source DB:  PubMed          Journal:  Can J Microbiol        ISSN: 0008-4166            Impact factor:   2.419


  1 in total

1.  Phoenix 100 versus Vitek 2 in the identification of gram-positive and gram-negative bacteria: a comprehensive meta-analysis.

Authors:  Kalliopi-Stavroula Chatzigeorgiou; Theodoros N Sergentanis; Sotirios Tsiodras; Stavros J Hamodrakas; Pantelis G Bagos
Journal:  J Clin Microbiol       Date:  2011-07-13       Impact factor: 5.948

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.