| Literature DB >> 20451431 |
Cedric Hugon1, Francesca D'Amico, Guy Aubert, Dimitris Sakellariou.
Abstract
Starting from general results of magnetostatics, we give fundamental considerations on the design and characterization of permanent magnets for NMR based on harmonic analysis and symmetry. We then propose a simple geometry that takes advantage of some of these considerations and discuss the practical aspects of the assembly of a real magnet based on this geometry, involving the characterization of its elements, the optimization of the layout and the correction of residual inhomogeneities due to material and geometry imperfections. We report with this low-cost, light-weight magnet (100 euros and 1.8 kg including the aluminum frame) a field of 120 mT (5.1 MHz proton) with a 10 ppm natural homogeneity over a sphere of 1.5 mm in diameter. Copyright (c) 2010 Elsevier Inc. All rights reserved.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20451431 DOI: 10.1016/j.jmr.2010.04.003