Automatic needle biopsy in the diagnosis of early breast cancer.

V Barreto1, H Hamed, A B Griffiths, A Hanby, M A Chaudary, I S Fentiman.   

Abstract

An automatic needle biopsy system, Bioptycut, has been evaluated for use in the outpatient biopsy of 107 patients with suspected operable breast cancer. Of the 107 patients, 96 proved to have malignancy. The sensitivity of the test was 65%, and the specificity 100%. Using a fine gauge needle the sensitivity was originally less than that of Trucut biopsy, but improved results were obtained with more experience.

Entities:  

Mesh:

Year:  1991        PMID: 2044776

Source DB:  PubMed          Journal:  Eur J Surg Oncol        ISSN: 0748-7983            Impact factor:   4.424