Accuracy of amplitude integrated EEG in a neonatal cohort.

E Evans1, S Koh, Jt Lerner, R Sankar, M Garg.   

Abstract

OBJECTIVE: To determine the sensitivity and specificity of amplitude integrated electroencephalogram (aEEG) compared to simultaneous standard electroencephalogram (sEEG) for seizure detection and background discontinuity.
DESIGN: Prospective paired cohort.
SETTING: Tertiary academic neonatal intensive care unit. PATIENTS: Infants were recruited from 2005 to 2008. Neonates requiring sEEG were recruited for simultaneous aEEG.
INTERVENTIONS: Following sEEG and aEEG, seizures were recorded as present or absent, and background was recorded as normal or discontinuous in each format. MAIN OUTCOME MEASURES: Presence of at least one seizure during recording. The background activity was reported as normal or discontinuous. Discontinuity of brain activity was further ranked as mild, moderate or severe.
RESULTS: 51 sEEG and aEEG studies were completed. 44 studies were analysed for presence of seizures and 46 were analysed for background discontinuity. Sensitivity for presence of seizures by aEEG was 80% and specificity was 50%. The proportion of infants with seizures were overdiagnosed by aEEG (63.6% vs 45.5% for sEEG p=0.045). Discontinuity of background activity had higher sensitivity (88.6%) and specificity (54.5%) when compared with seizure detection. When stratified by indication for EEG, hypoxic episode (n=14) or suspected seizures (n=33), similar sensitivity for presence of seizure (80%) was noted by aEEG and sEEG. However the specificity of aEEG for seizure detection was higher in neonates undergoing EEG for suspected seizures (66.7% vs 22.2%).
CONCLUSIONS: Background abnormalities were detected with fair accuracy by aEEG but aEEG criteria alone would result in the overdiagnosis of neonatal seizures. Therefore seizures noted on aEEG require sEEG confirmation prior to implementing anticonvulsant therapy for neonatal seizures.

Entities:  

Mesh:

Year:  2010        PMID: 20444809     DOI: 10.1136/adc.2009.165969

Source DB:  PubMed          Journal:  Arch Dis Child Fetal Neonatal Ed        ISSN: 1359-2998            Impact factor:   5.747


  13 in total

Review 1.  Neurodiagnostic techniques in neonatal critical care.

Authors:  Taeun Chang; Adre du Plessis
Journal:  Curr Neurol Neurosci Rep       Date:  2012-04       Impact factor: 5.081

Review 2.  Neonatal seizures and status epilepticus.

Authors:  Nicholas S Abend; Courtney J Wusthoff
Journal:  J Clin Neurophysiol       Date:  2012-10       Impact factor: 2.177

3.  Seizures in Preterm Neonates: A Multicenter Observational Cohort Study.

Authors:  Hannah C Glass; Renée A Shellhaas; Tammy N Tsuchida; Taeun Chang; Courtney J Wusthoff; Catherine J Chu; M Roberta Cilio; Sonia L Bonifacio; Shavonne L Massey; Nicholas S Abend; Janet S Soul
Journal:  Pediatr Neurol       Date:  2017-04-20       Impact factor: 3.372

4.  Diagnostic value of amplitude-integrated electroencephalogram in neonatal seizures.

Authors:  Lian Zhang; Yan-Xia Zhou; Li-Wen Chang; Xiao-Ping Luo
Journal:  Neurosci Bull       Date:  2011-08       Impact factor: 5.203

5.  Seizure Detection by Critical Care Providers Using Amplitude-Integrated Electroencephalography and Color Density Spectral Array in Pediatric Cardiac Arrest Patients.

Authors:  Geneviève Du Pont-Thibodeau; Sarah M Sanchez; Abbas F Jawad; Vinay M Nadkarni; Robert A Berg; Nicholas S Abend; Alexis A Topjian
Journal:  Pediatr Crit Care Med       Date:  2017-04       Impact factor: 3.624

Review 6.  Electroencephalographic monitoring in the pediatric intensive care unit.

Authors:  Nicholas S Abend; Kevin E Chapman; William B Gallentine; Joshua Goldstein; Ann E Hyslop; Tobias Loddenkemper; Kendall B Nash; James J Riviello; Cecil D Hahn
Journal:  Curr Neurol Neurosci Rep       Date:  2013-03       Impact factor: 5.081

Review 7.  Amplitude-integrated electro-encephalography: the child neurologist's perspective.

Authors:  Hannah C Glass; Courtney J Wusthoff; Renée A Shellhaas
Journal:  J Child Neurol       Date:  2013-05-20       Impact factor: 1.987

Review 8.  Lost in Transition: A Systematic Review of Neonatal Electroencephalography in the Delivery Room-Are We Forgetting an Important Biomarker for Newborn Brain Health?

Authors:  Daragh Finn; Eugene M Dempsey; Geraldine B Boylan
Journal:  Front Pediatr       Date:  2017-08-10       Impact factor: 3.418

Review 9.  A Comparison Of Conventional Electroencephalography With Amplitude-Integrated EEG In Detection Of Neonatal Seizures.

Authors:  Maliheh Kadivar; Elahe Movahedi Moghadam; Reza Shervin Badv; Raziye Sangsari; Maryam Saeedy
Journal:  Med Devices (Auckl)       Date:  2019-12-10

Review 10.  Enhanced Monitoring of the Preterm Infant during Stabilization in the Delivery Room.

Authors:  Daragh Finn; Geraldine B Boylan; C Anthony Ryan; Eugene M Dempsey
Journal:  Front Pediatr       Date:  2016-03-31       Impact factor: 3.418

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.