Literature DB >> 20442358

Development of a novel prohealing stent designed to deliver sirolimus from a biodegradable abluminal matrix.

Juan F Granada1, Shigenobu Inami, Michael S Aboodi, Armando Tellez, Krzysztof Milewski, David Wallace-Bradley, Sherry Parker, Steve Rowland, Gaku Nakazawa, Marc Vorpahl, Frank D Kolodgie, Greg L Kaluza, Martin B Leon, Renu Virmani.   

Abstract

BACKGROUND: We aimed to demonstrate that, by separating endothelial progenitor cell capture from sirolimus delivery through the application of drug to the abluminal surface of the stent, the degree of endothelialization can be enhanced. METHODS AND
RESULTS: Stainless steel R Stents, with biodegradable SynBiosys polymer coating with sirolimus abluminally applied and surface modified with anti-CD34 antibody were prepared at 2 dosages (low-dose sirolimus [LD-Combo, 2.5 microg sirolimus/mm] and full-dose sirolimus [Combo, 5 microg sirolimus/mm). These Combo stents and the Cypher stent (10 microg sirolimus/mm) were deployed in 98 normal porcine arteries and harvested for pharmacokinetic analysis at 0.25, 1, 3, 7, 14, 28, and 35 days. The LD-Combo stents showed faster early release (50%total dose in 72 hours) than the Combo and Cypher. At 30 days, drug release was near complete with both Combo stents, whereas 20% of drug remained on the Cypher stents. To assess efficacy, a total of 50 stents (Xience V=8, Cypher=8, Genous bioengineered R stent=6, LD-Combo=14, and Combo=14) were implanted in 18 pigs for 14 and 28 days. Optical coherence tomography was performed, and stents were harvested for histology. At 28 days, there was less neointimal thickness with Combo (0.173+/-0.088 mm) compared with Cypher (0.358+/-0.225 mm), LD-Combo (0.316+/-0.228 mm), and Xience V (0.305+/-0.252 mm; P<0.00001). Immunohistochemical analysis of endothelialization showed that Genous bioengineered R stent had the highest degree of platelet endothelial cell adhesion molecule expression (87%) followed by the Combo (75%), LD-Combo (65%), and Cypher (58%).
CONCLUSIONS: Both optical coherence tomography and histology demonstrate that anti-CD34 sirolimus-eluting stents promote endothelialization while reducing neointimal formation and inflammation.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20442358     DOI: 10.1161/CIRCINTERVENTIONS.109.919936

Source DB:  PubMed          Journal:  Circ Cardiovasc Interv        ISSN: 1941-7640            Impact factor:   6.546


  18 in total

Review 1.  Restenosis after PCI. Part 2: prevention and therapy.

Authors:  J Wouter Jukema; Tarek A N Ahmed; Jeffrey J W Verschuren; Paul H A Quax
Journal:  Nat Rev Cardiol       Date:  2011-10-11       Impact factor: 32.419

2.  Accelerated endothelialization with a polymer-free sirolimus-eluting antibody-coated stent.

Authors:  Zhanjiang Yu; Huagang Zhu; Shuzheng Lü; Xiaoda Yang
Journal:  J Mater Sci Mater Med       Date:  2013-08-23       Impact factor: 3.896

3.  Prohealing endothelial progenitor cell capture stents: do the cells captured explain the clinical outcomes?

Authors:  Jane A Leopold
Journal:  Circ Cardiovasc Interv       Date:  2013-10-01       Impact factor: 6.546

4.  Assessment of an asymmetrical coating stent with sirolimus released from ablumial matrix in porcine model.

Authors:  Li Shen; Yizhe Wu; Feng Zhang; Lingxiang Wu; Chao Dong; Yongxing Gao; Aijun Sun; Yunzeng Zou; Juying Qian; Jianguo Sun; Wei Zhong; Junbo Ge
Journal:  Clin Res Cardiol       Date:  2012-05-25       Impact factor: 5.460

5.  Direct Adsorption of Anti-CD34 Antibodies on the Nano-Porous Stent Surface to Enhance Endothelialization.

Authors:  Guowei Fu; Zhanjiang Yu; Yongqiang Chen; Yundai Chen; Feng Tian; Xiaoda Yang
Journal:  Acta Cardiol Sin       Date:  2016-05       Impact factor: 2.672

6.  Enhanced endothelialization of a new stent polymer through surface enhancement and incorporation of growth factor-delivering microparticles.

Authors:  Hao Xu; Kytai T Nguyen; Emmanouil S Brilakis; Jian Yang; Eric Fuh; Subhash Banerjee
Journal:  J Cardiovasc Transl Res       Date:  2012-05-26       Impact factor: 4.132

Review 7.  Concise Review: Endothelial Progenitor Cells in Regenerative Medicine: Applications and Challenges.

Authors:  Mark Seow Khoon Chong; Wei Kai Ng; Jerry Kok Yen Chan
Journal:  Stem Cells Transl Med       Date:  2016-03-08       Impact factor: 6.940

8.  Preclinical evaluation of a novel abluminal surface coated sirolimus eluting stent with biodegradable polymer matrix.

Authors:  Prakash Sojitra; Manish Doshi; Marco Galloni; Christina Vignolini; Ashwin Vyas; Bhavesh Chevli; Imad Sheiban
Journal:  Cardiovasc Diagn Ther       Date:  2015-08

Review 9.  Neoatherosclerosis: Coronary stents seal atherosclerotic lesions but result in making a new problem of atherosclerosis.

Authors:  Hidenori Komiyama; Masamichi Takano; Noritake Hata; Yoshihiko Seino; Wataru Shimizu; Kyoichi Mizuno
Journal:  World J Cardiol       Date:  2015-11-26

10.  Automatic stent detection in intravascular OCT images using bagged decision trees.

Authors:  Hong Lu; Madhusudhana Gargesha; Zhao Wang; Daniel Chamie; Guilherme F Attizzani; Tomoaki Kanaya; Soumya Ray; Marco A Costa; Andrew M Rollins; Hiram G Bezerra; David L Wilson
Journal:  Biomed Opt Express       Date:  2012-10-15       Impact factor: 3.732

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.