Literature DB >> 2043723

Demonstration of the compatibility of narrow-bore packed column high-performance liquid chromatography with conventional detection systems.

D N Mallett1, B Law.   

Abstract

A number of commercially available high-performance liquid chromatography detectors fitted with conventionally sized flow cells (4.5 to 14 microliters) have been evaluated for use for narrow-bore (1-3 mm i.d.) packed columns. Contrary to the popular misconception, small flow cell volume (and consequently short path length) does not appear to be a prerequisite for successful use of narrow-bore columns. The 1 mm i.d. columns when used with certain detectors gave up to a 17-fold increase in sensitivity, compared with conventional 4.6 mm i.d. columns. This work, which was carried out using conventional injection volumes (50 microliters), shows that narrow-bore HPLC columns can be used in most laboratories to significant advantage without costly conversion of existing equipment.

Mesh:

Year:  1991        PMID: 2043723     DOI: 10.1016/0731-7085(91)80237-4

Source DB:  PubMed          Journal:  J Pharm Biomed Anal        ISSN: 0731-7085            Impact factor:   3.935


  2 in total

1.  Measurement of urinary catecholamines in small samples for mice.

Authors:  Emily A Smith; Abigail L Schwartz; James B Lucot
Journal:  J Pharmacol Toxicol Methods       Date:  2012-09-07       Impact factor: 1.950

2.  Diclofenac and naproxen analysis by microbore liquid chromatography (LC) with native fluorescence detection.

Authors:  W R Baeyens; G van der Weken; M Schelkens
Journal:  J Fluoresc       Date:  1995-06       Impact factor: 2.217

  2 in total