Literature DB >> 20436002

Long-term survival of the Souter-Strathclyde total elbow replacement in patients with rheumatoid arthritis.

M Ikävalko1, R Tiihonen, E T Skyttä, E A Belt.   

Abstract

Between 1982 and 1997, 403 consecutive patients (522 elbows) with rheumatoid arthritis underwent Souter-Strathclyde total elbow replacement. By the end of 2007, there had been 66 revisions for aseptic loosening in 60 patients. The mean time of follow-up was 10.6 years (0 to 25) The survival rates at five-, ten, 15 and 19 years were 96% (95%, confidence interval (CI) 95 to 98), 89% (95% CI 86 to 92), 83% (95% CI 78 to 87), and 77% (95% CI 69 to 85), respectively. The small and medium-sized short-stemmed primary humeral components had a 5.6-fold and 3.6-fold risk of revision for aseptic loosening respectively, compared to the medium-sized long-stemmed component. The small and medium-sized all-polyethylene ulnar components had respectively a 28.2-fold and 8.4-fold risk of revision for aseptic loosening, compared to the metal-backed ulnar components. The use of retentive ulnar components was not associated with an increased risk of aseptic loosening compared to non-retentive implants.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20436002     DOI: 10.1302/0301-620X.92B5.22613

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  3 in total

Review 1.  What design and material factors impact the wear and corrosion performance in total elbow arthroplasties?

Authors:  Mark P Figgie; Timothy M Wright; Denise Drinkwater
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

2.  Primary Linked Semiconstrained Total Elbow Arthroplasty for Rheumatoid Arthritis: A Single-Institution Experience with 461 Elbows Over Three Decades.

Authors:  Joaquin Sanchez-Sotelo; Yaser M K Baghdadi; Bernard F Morrey
Journal:  J Bone Joint Surg Am       Date:  2016-10-19       Impact factor: 5.284

3.  RESULTS FROM BI-CONTACT(®) TOTAL ELBOW ARTHROPLASTY: MULTICENTER STUDY.

Authors:  Eduardo Benegas; Eduardo Angeli Malavolta; Mauro Emilio Conforto Gracitelli; Augusto Tadeu Barros de Sousa; Alberto Naoki Miyazaki; Marcelo Fregoneze; Roberto Yukio Ikemoto; Joel Murachovsky; Marcelo Hide Matsumoto; Marcel Jun Sugawara Tamaoki; Arnaldo Amado Ferreira Neto
Journal:  Rev Bras Ortop       Date:  2015-12-06
  3 in total