Literature DB >> 20435203

High incidence of thrombus formation at 18 months after paclitaxel-eluting stent implantation: angioscopic comparison with sirolimus-eluting stent.

Masahiko Hara1, Masami Nishino, Masayuki Taniike, Nobuhiko Makino, Hiroyasu Kato, Yasuyuki Egami, Ryu Shutta, Hitoshi Yamaguchi, Jun Tanouchi, Yoshio Yamada.   

Abstract

BACKGROUND: Difference of neointimal formational pattern and incidence of thrombus formation between sirolimus-eluting stent (SES) and paclitaxel-eluting stent (PES) at 18 months after stent implantation has not previously been reported.
METHODS: We retrospectively enrolled 35 event-free patients who received SES (15 patients, 18 stents) or PES (20 patients, 23 stents) implantation with 18-month follow-up coronary angioscopy. We divided our patients into SES or PES groups and compared neointimal coverage pattern and incidence of red mural thrombus formation between the 2 groups. Neointimal coverage grades were classified into 4 categories. Minimum neointimal coverage grade, maximum grade, and heterogeneity score were assessed in each stent. Heterogeneity score was calculated by subtracting minimum from maximum grade within one stent.
RESULTS: Minimum neointimal coverage grade of PES was significantly lower than that of SES (0.70 +/- 0.64 vs 1.33 +/- 0.69, P = .005), whereas maximum grade was not significantly different (2.48 +/- 0.73 vs 2.22 +/- 0.73, P = .218). Heterogeneity score and incidence of red mural thrombus of PES were higher than those of SES (1.78 +/- 0.80 vs 0.89 +/- 0.76, P = .002 and 70% vs 11%, P < .001).
CONCLUSIONS: The present study revealed that PES shows more heterogeneous neointimal coverage and higher incidence of thrombus formation as compared with SES at 18 months after stent implantation. 2010 Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20435203     DOI: 10.1016/j.ahj.2010.02.032

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  2 in total

1.  A higher colour grade yellow plaque was detected at one year after implantation of an everolimus-eluting stent than after a zotarolimus-eluting stent.

Authors:  Koshi Matsuo; Yasunori Ueda; Mayu Nishio; Akio Hirata; Mitsutoshi Asai; Takayoshi Nemoto; Ayaka Murakami; Kazunori Kashiwase; Kazuhisa Kodama
Journal:  Heart Asia       Date:  2013-09-07

2.  Serial angioscopic evaluation of neointimal coverage and incidence of thrombus formation after Paclitaxel-eluting stent implantation: comparison between 6- and 18-month follow-up.

Authors:  Masahiko Hara; Masami Nishino; Masayuki Taniike; Nobuhiko Makino; Hiroyasu Kato; Yasuyuki Egami; Ryu Shutta; Jun Tanouchi; Yoshio Yamada
Journal:  Clin Cardiol       Date:  2011-03-22       Impact factor: 2.882

  2 in total