Literature DB >> 20426636

Radiofrequency perforation system for in vivo antegrade fenestration of aortic stent-grafts.

Leonard W H Tse1, Sophie Lerouge, Bao T Bui, Eric Therasse, Hélène Héon, Gilles Soulez.   

Abstract

PURPOSE: To evaluate the ability to create in vivo antegrade fenestrations (IVAF) using a radiofrequency probe in juxtarenal aortic stent-grafts to preserve the patency of renal arteries.
METHODS: Modified stent-grafts with an unsupported fenestration area were deployed in the juxtarenal aorta of 2 25-kg dogs. Prior to deployment, both renal arteries were marked with detachable coils that were later removed (dog 1) or with hydrophilic catheters (dog 2). After deployment, a radiofrequency probe was used to perforate the graft. If puncture was successful, attempts were made to pass guidewires into the renal arteries for balloon dilation and stenting of the graft perforation. Doppler ultrasound scans of the renal arteries were taken after 1 week. Doppler ultrasound and angiography were repeated prior to autopsy and macroscopic analysis at 1 month.
RESULTS: Fenestration was attempted in 3 renal arteries (1 in dog 1 and 2 in dog 2). It was successful in 2 renal arteries but required 6 attempts in dog 1 and 3 in dog 2. The failed case was related to the position of a stent-graft strut close to the fenestration area. One stent thrombosis was seen during follow-up, while the other fenestrated artery remained patent. No stent fracture was encountered during follow-up.
CONCLUSION: This radiofrequency device allowed successful fenestration but was difficult to control. Technical improvements are required to improve clinical success and patency.

Entities:  

Mesh:

Year:  2010        PMID: 20426636     DOI: 10.1583/09-2903.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  3 in total

1.  Aortic arch aneurysm repair using the Najuta stent graft in a challenging compromised seal zone.

Authors:  Naoki Toya; Kota Shukuzawa; Soichiro Fukushima; Masamichi Momose; Tadashi Akiba; Takao Ohki
Journal:  J Vasc Surg Cases       Date:  2016-02-23

2.  Antegrade in situ laser fenestration of aortic stent graft during endovascular aortic repair: A case report.

Authors:  Zhi-Wei Wang; Zhen-Tao Qiao; Ming-Xing Li; Hua-Long Bai; Yuan-Feng Liu; Tao Bai
Journal:  World J Clin Cases       Date:  2022-02-06       Impact factor: 1.337

3.  Evaluation for the safety and effectiveness of the in situ fenestration system in TEVAR for aortic arch pathologies: protocol for a prospective, multicentre and single-arm study.

Authors:  Xiaofan Sun; Yuanqing Kan; Lihong Huang; Zhihui Dong; Daqiao Guo; Yi Si; Weiguo Fu
Journal:  BMJ Open       Date:  2021-02-22       Impact factor: 2.692

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.