Accuracy of three transcutaneous carbon dioxide monitors in critically ill children.

Javier Urbano1, Verónica Cruzado, Jesús López-Herce, Jimena del Castillo, José María Bellón, Angel Carrillo.   

Abstract

OBJECTIVES: To study the accuracy of three devices for measuring transcutaneous CO(2) tension in critically ill children.
METHODS: A prospective study comparing the values from three transcutaneous CO(2) monitors (SenTec, TOSCA 500, and TINA TCM3) with simultaneous arterial CO(2) (PaCO(2)) and end-tidal CO(2) (EtCO(2)) values. Clinical data were collected from the patients. Influence of core-skin temperature gradient and doses of inotropic drugs was evaluated.
RESULTS: There were 62 samples from 41 critically ill children with ages between 2 and 192 months (median, 18.5 months) and weights between 3.1 and 72 kg (median, 9 kg). The median PaCO(2) was 42.5 mmHg (range, 28-85 mmHg). Transcutaneous CO(2) (PtcCO(2)) values correlated better with PaCO(2) than with EtCO(2). The correlation coefficient between PaCO(2) and PtCO(2) was 0.833 with the TINA TCM3 monitor, 0.931 with the SenTec monitor, and 0.765 with the TOSCA 500 monitor. The mean (SD) differences between the PaCO(2) and PtcCO(2) were 4.5 (3.7) mmHg, 4.3 (3.8) mmHg, and 5.6 (5.1) mmHg, respectively, with the three monitors, and the differences between the PaCO(2) and PtcCO(2) were less than 7.5 mmHg in 77.7%, 81.2%, and 67.7% of the samples. Bland-Altman analysis showed a precision of +/-11.5 mmHg for TINA TCM3 monitor, +/-10.6 mmHg for SenTec monitor, and +/-14.8 mmHg for TOSCA monitor. No influence of core-skin temperature gradient and inotropic index on the differences between PaCO(2) and PtcCO(2) was observed.
CONCLUSIONS: The three transcutaneous CO(2) monitors have an acceptable correlation with arterial CO(2) tension and can be useful in critically ill children.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20425856     DOI: 10.1002/ppul.21203

Source DB:  PubMed          Journal:  Pediatr Pulmonol        ISSN: 1099-0496


  4 in total

1.  Transcutaneous PCO2 monitoring in infants hospitalized with viral bronchiolitis.

Authors:  S Gal; A Riskin; I Chistyakov; N Shifman; I Srugo; A Kugelman
Journal:  Eur J Pediatr       Date:  2014-08-28       Impact factor: 3.183

2.  Evaluation of a transcutaneous carbon dioxide monitor in patients with acute respiratory failure.

Authors:  Antonello Nicolini; Maura Bravo Ferrari
Journal:  Ann Thorac Med       Date:  2011-10       Impact factor: 2.219

3.  End-tidal carbon dioxide measurement in preterm infants with low birth weight.

Authors:  Hsin-Ju Lin; Ching-Tzu Huang; Hsiu-Feng Hsiao; Ming-Chou Chiang; Mei-Jy Jeng
Journal:  PLoS One       Date:  2017-10-17       Impact factor: 3.240

4.  Comparison of arterial CO2 estimation by end-tidal and transcutaneous CO2 measurements in intubated children and variability with subject related factors.

Authors:  Muhterem Duyu; Yasemin Mocan Çağlar; Zeynep Karakaya; Mine Usta Aslan; Seyhan Yılmaz; Aslı Nur Ören Leblebici; Anıl Doğan Bektaş; Meral Bahar; Meryem Nihal Yersel
Journal:  J Clin Monit Comput       Date:  2020-07-27       Impact factor: 2.502

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.