A new technique to assist epidural needle placement: fiberoptic-guided insertion using two wavelengths.

Chien-Kun Ting1, Mei-Yung Tsou, Pin-Tarng Chen, Kuang-Yi Chang, M Susan Mandell, Kwok-Hon Chan, Yin Chang.   

Abstract

BACKGROUND: Up to 10% of epidurals fail due to incorrect catheter placement. We describe a novel optical method to assist epidural catheter insertion in a porcine model.
METHODS: Optical emissions were tested on ex vivo tissues from porcine paravertebral tissues to identify optical reflective spectra. The wavelengths of 650 and 532 nm differentiated epidural space from the ligamentum flavum. We then used a hollow stylet that contained optical fibers to place epidural needles in anesthetized pigs. Real-time data were displayed on an oscilloscope and stored for analysis. A total of 50 punctures were done in four laboratory pigs. Data were expressed as mean +/- SD.
RESULTS: Paired t test shows significant optical differences between the epidural space and the ligamentum flavum at both 650 nm (P < 0.001) and 532 nm (P = 0.014). Mean magnitudes for 650 nm, 532 nm, and their ratio were 3.565 +/- 0.194, 2.542 +/- 0.145, and 0.958 +/- 0.172 at epidural space and 3.842 +/- 0.191, 2.563 +/- 0.131, and 1.228 +/- 0.244 at ligamentum flavum, respectively. There were no differences in the optical characteristics of the ligamentum flavum and epidural space at different levels in the lumbar and thoracic region (two-way ANOVA P > 0.05).
CONCLUSIONS: This is the first study to introduce a new optical method to localize epidural space in a porcine model. Epidural space could be identified by the changes in the reflective pattern of light emitted at 650 nm, which were specific for the ligamentum flavum and dural tissue. Real-time optical information successfully guided a modified Tuohy needle into the epidural space.

Entities:  

Mesh:

Year:  2010        PMID: 20418693     DOI: 10.1097/ALN.0b013e3181d3d958

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  14 in total

1.  Eyes in the needle: novel epidural needle with embedded high-frequency ultrasound transducer--epidural access in porcine model.

Authors:  Huihua K Chiang; Qifa Zhou; M Susan Mandell; Mei-Yung Tsou; Shih-Pin Lin; K Kirk Shung; Chien-Kun Ting
Journal:  Anesthesiology       Date:  2011-06       Impact factor: 7.892

2.  Epidural catheter with integrated light guides for spectroscopic tissue characterization.

Authors:  R P Soto-Astorga; S West; S Putnis; J C Hebden; A E Desjardins
Journal:  Biomed Opt Express       Date:  2013-10-24       Impact factor: 3.732

Review 3.  [Seeing more : Technical innovations in regional anesthesia].

Authors:  T Wiesmann; T Steinfeldt; T Volk; U Schwemmer; P Kessler; H Wulf
Journal:  Anaesthesist       Date:  2014-11       Impact factor: 1.041

4.  Real-time epidural anesthesia guidance using optical coherence tomography needle probe.

Authors:  Qinggong Tang; Chia-Pin Liang; Kyle Wu; Anthony Sandler; Yu Chen
Journal:  Quant Imaging Med Surg       Date:  2015-02

5.  Intelligent epidural needle placement using fiber-probe optical coherence tomography in a piglet model.

Authors:  Meng-Chun Kao; Yu-Te Wu; Mei-Yung Tsou; Wen-Chuan Kuo; Chien-Kun Ting
Journal:  Biomed Opt Express       Date:  2018-07-13       Impact factor: 3.732

6.  Raman Spectroscopy Differentiates Each Tissue from the Skin to the Spinal Cord: A Novel Method for Epidural Needle Placement?

Authors:  T Anthony Anderson; Jeon Woong Kang; Tatyana Gubin; Ramachandra R Dasari; Peter T C So
Journal:  Anesthesiology       Date:  2016-10       Impact factor: 7.892

7.  Portable optical epidural needle-a CMOS-based system solution and its circuit design.

Authors:  Cihun-Siyong Alex Gong; Shih-Pin Lin; M Susan Mandell; Mei-Yung Tsou; Yin Chang; Chien-Kun Ting
Journal:  PLoS One       Date:  2014-08-27       Impact factor: 3.240

8.  Epidural needle with embedded optical fibers for spectroscopic differentiation of tissue: ex vivo feasibility study.

Authors:  Adrien E Desjardins; Benno H W Hendriks; Marjolein van der Voort; Rami Nachabé; Walter Bierhoff; Guus Braun; Drazenko Babic; James P Rathmell; Staffan Holmin; Michael Söderman; Björn Holmström
Journal:  Biomed Opt Express       Date:  2011-05-06       Impact factor: 3.732

9.  An alternative management procedure after inadvertent dural puncture.

Authors:  Kemal Tolga Saracoglu; Ayten Saracoglu; Filiz Karaca; Vural Fidan
Journal:  J Res Med Sci       Date:  2011-03       Impact factor: 1.852

10.  Investigation of an intelligent system for fiber optic-based epidural anesthesia.

Authors:  Cihun-Siyong Alex Gong; Chien-Kun Ting
Journal:  Biomed Res Int       Date:  2014-03-20       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.