Literature DB >> 20417999

Tightening force and torque of nonlocking screws in a reverse shoulder prosthesis.

A Terrier1, S H Kochbeck, F Merlini, M Gortchacow, D P Pioletti, A Farron.   

Abstract

BACKGROUND: Reversed shoulder arthroplasty is an accepted treatment for glenohumeral arthritis associated to rotator cuff deficiency. For most reversed shoulder prostheses, the baseplate of the glenoid component is uncemented and its primary stability is provided by a central peg and peripheral screws. Because of the importance of the primary stability for a good osteo-integration of the baseplate, the optimal fixation of the screws is crucial. In particular, the amplitude of the tightening force of the nonlocking screws is clearly associated to this stability. Since this force is unknown, it is currently not accounted for in experimental or numerical analyses. Thus, the primary goal of this work is to measure this tightening force experimentally. In addition, the tightening torque was also measured, to estimate an optimal surgical value.
METHODS: An experimental setup with an instrumented baseplate was developed to measure simultaneously the tightening force, tightening torque and screwing angle, of the nonlocking screws of the Aquealis reversed prosthesis. In addition, the amount of bone volume around each screw was measured with a micro-CT. Measurements were performed on 6 human cadaveric scapulae.
FINDINGS: A statistically correlated relationship (p<0.05, R=0.83) was obtained between the maximal tightening force and the bone volume. The relationship between the tightening torque and the bone volume was not statistically significant.
INTERPRETATION: The experimental relationship presented in this paper can be used in numerical analyses to improve the baseplate fixation in the glenoid bone. Copyright (c) 2010 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20417999     DOI: 10.1016/j.clinbiomech.2010.03.011

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  3 in total

1.  Biomechanical evaluation of reconstruction plates with locking, nonlocking, and hybrid screws configurations in calcaneal fracture: a finite element model study.

Authors:  Ching-Hsuan Chen; Chinghua Hung; Yu-Chun Hsu; Chen-Sheng Chen; Chao-Ching Chiang
Journal:  Med Biol Eng Comput       Date:  2017-02-21       Impact factor: 2.602

2.  Fixation stability comparison of bone screws based on thread design: buttress thread, triangle thread, and square thread.

Authors:  Xiaoreng Feng; Zhaopei Luo; Yupeng Li; Yiyi Yao; Weichen Qi; Bin Chen; Hongfeng Liang
Journal:  BMC Musculoskelet Disord       Date:  2022-08-30       Impact factor: 2.562

3.  CT scan evaluation of glenoid bone and pectoralis major tendon: interest in shoulder prosthesis.

Authors:  Laurent Obert; Christelle Peyron; Etienne Boyer; Gauthier Menu; François Loisel; Sébastien Aubry
Journal:  SICOT J       Date:  2016-10-07