Literature DB >> 20416521

Accuracy of a laboratory-based computer implant guiding system.

Eitan Barnea1, Ido Alt, Roni Kolerman, Joseph Nissan.   

Abstract

OBJECTIVE: Computer-guided implant placement is a growing treatment modality in partially and totally edentulous patients, though data about the accuracy of some systems for computer-guided surgery is limited. The purpose of this study was to evaluate the accuracy of a laboratory computer-guided system. STUDY
DESIGN: A laboratory-based computer guiding system (M Guide; MIS technologies, Shlomi, Israel) was used to place implants in a fresh sheep mandible. A second computerized tomography (CT) scan was taken after placing the implants . The drill plan figures of the planned implants were positioned using assigned software (Med3D, Heidelberg, Germany) on the second CT scan to compare the implant position with the initial planning. Values representing the implant locations of the original drill plan were compared with that of the placed implants using SPSS software.
RESULTS: Six measurements (3 vertical, 3 horizontal) were made on each implant to assess the deviation from the initial implant planning. A repeated-measurement analysis of variance was performed comparing the location of measurement (center, abutment, apex) and type of deviation (vertical vs. horizontal). The vertical deviation (mean -0.168) was significantly smaller than the horizontal deviation (mean 1.148).
CONCLUSION: The laboratory computer-based guiding system may be a viable treatment concept for placing implants. Copyright (c) 2010 Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20416521     DOI: 10.1016/j.tripleo.2010.01.001

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  3 in total

1.  Implant Bed Preparation with an Erbium, Chromium Doped Yttrium Scandium Gallium Garnet (Er,Cr: YSGG) Laser Using Stereolithographic Surgical Guide.

Authors:  Gülin Seymen; Zeynep Turgut; Gizem Berk; Ayşen Bodur
Journal:  J Lasers Med Sci       Date:  2013

2.  Accuracy of computer-assisted surgery in immediate implant placement: An experimental study.

Authors:  Nicole Báez-Marrero; José Luis Rafel; Yalil Augusto Rodríguez-Cárdenas; Aron Aliaga-Del Castillo; Heraldo Luis Dias-Da Silveira; Luis Ernesto Arriola-Guillén
Journal:  J Indian Soc Periodontol       Date:  2021-09-27

3.  Accuracy of computer-assisted implant placement with insertion templates.

Authors:  Eleni Naziri; Alexander Schramm; Frank Wilde
Journal:  GMS Interdiscip Plast Reconstr Surg DGPW       Date:  2016-05-13
  3 in total